Mounting and installation
Adjustment elements
Please note!
5

Setup

5.1

General information

Take note of the mounting and installation instructions in the operating
instructions manual.
Setup must be carried out under process conditions.
5.2

Adjustment instructions

The adjustment elements must be set according to the specified
safety function:
Slide switch for changeover of the mode (min./max.)
Slide switch for changeover of the sensitivity
The function of the adjustment elements is described in the operating
instructions manual.
During adjustment process, the safety function must be considered
as unreliable!
If necessary, you must take other measures to maintain the safety
function.
With regard to the switch on/swich off delay it must be ensured that
the sum of all switching delays from the transducer to the actuator is
adapted to the process safety time!
The instrument must be protected against inadvertent or unauthorized
adjustment!
SITRANS LVL200H - Safety Manual