Operating The Alcohol Wet Bath Simulator - PAS Systems International PAS V Calibration Manual

Passive alcohol sensor
Hide thumbs Also See for PAS V:
Table of Contents

Advertisement

Operating the Alcohol Wet Bath Simulator

On/Off Switch
Outlet to PAS V
The Alcohol Wet Bath Simulator
The lower (jar) portion is filled with a heated solution of water and alcohol, which
simulates the blood running through the arteries in the lungs. The upper portion (the
head space) is filled with a mixture of air and alcohol that simulates the air in the lungs
that is in contact with the blood across the cell membranes. Air flowing out of the outlet
pipe simulates the expired air of a person who has been drinking.
In order to work properly, the temperature of the water/alcohol solution must be 34° C
(body temperature) and the mixture of alcohol and water must be accurate for the BrAC
to be used in the test (.01 or .005 BrAC).
Preparing Wet Bath Simulator
1.
Remove the glass jar from the simulator by unscrewing it from the top assembly.
Caution: DO NOT remove Top Assembly and expose the heating element to open
air with power turned ON. This will result in damage to the heating element.
2.
Place 500 ml of the .005 calibration solution into the simulator.
3.
Assemble by tightly turning the jar back into the top assembly.
4.
Add tubing to the inlet and outlet of the simulator.
5.
Check the seal of the lid on the jar by blowing into the inlet while holding outlet
closed. There should be resistance to the airflow.
6.
Plug in the simulator and turn the power ON and allow the solution to heat to
34◦C (approx 15 min).
heating. The heater lamp goes off when the simulator fluid has reached 34◦C.
However, it will come back on from time to time to keep the fluid at 34◦C. When
34°C is reached, the simulator is ready to be used.
The heater lamp is lit when the heating element is
3
Inlet Tube
Heater ON Light