Equipment Checklist For Wet Bath Sensitivity Check/Calibration - PAS Systems International PAS V Calibration Manual

Passive alcohol sensor
Hide thumbs Also See for PAS V:
Table of Contents

Advertisement

PAS Systems International recommends having the calibration checked
once a year, or whenever it seems to be losing sensitivity. Instructions for
both wet bath and dry gas checking/calibrating are included in this manual.
You may also choose to send your PAS V into the manufacturer (PAS
Systems) to be calibrated.
Equipment Checklist for Wet Bath
Sensitivity Check/Calibration
The PAS V Passive Alcohol Sensor Flashlight
The "Calibration Assembly" (A modified plastic mouthpiece that
connects the outlet of the wet bath simulator to the intake port of the
PAS V. Available from PAS Systems).
A NHTSA approved Wet Bath Simulator
500 milliliters of .005% ethanol alcohol
lower, a 0.01% ethanol solution may be used.
Plastic tubing to connect the simulator's inlet to a mouthpiece into
which the operator blows (or can also be used to connect the
simulator's inlet to a wet bath simulator air pump).
A jeweler's screwdriver to adjust the fuel-cell sensitivity if calibration
should be necessary.
A PAS Cal-Pump wet bath simulator air pump (recommended)
*
Premixed Certified Stock Solutions are available from PAS Systems
International
solution.*