Omega CDH-SD11 User Manual

Omega CDH-SD11 User Manual

Conductivity, tds, and salinity meter with real time sd card data logger

Advertisement

Quick Links

2
TM
Conductivity, TDS, and Salinity Meter with Real Time SD Card Data Logger
User's Guide
Shop online at
omega.com
e
e-mail: info@omega.com
Fo
For latest product manuals:
www.omegamanual.info
CDH-SD11

Advertisement

Table of Contents
loading

Summary of Contents for Omega CDH-SD11

  • Page 1 User’s Guide Shop online at omega.com e-mail: info@omega.com For latest product manuals: www.omegamanual.info CDH-SD11 Conductivity, TDS, and Salinity Meter with Real Time SD Card Data Logger...
  • Page 2 Fax: (203) 359-7700 e-mail: info@omega.com For Other Locations Visit omega.com/worldwide The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3 8G SD card ADDITIONAL PURCHASE  CDE-11G K=1 electrode replacement (Graphite type)  CDE-11P K=10 option (Platinum type) SPECIFICATION CDH-SD11 Model Cond/EC/TDS /Salt Parameter Auto 2 secs, 5 secs, 10 secs, 15 secs, 30 secs, 60 secs, 120 Data logger sampling...
  • Page 4: Keypad (Controls)

    Sensor type Graphite plate TDS (Total Dissolved Solids) 0 to 19.99 ppm 0 to 199.9 ppm 0 to 1999 ppm Measurement range 0 to 19.99 ppt 0 to 199.9 ppt Accuracy ±1% of full scale + 1 digit Resolution 0.01 ppm, 0.1 ppm, 1 ppm ; 0.01 ppt, 0.1 ppt Salt Measurement range 0~42.0 ppt (Seawater)
  • Page 5 For short time measurement, you may power on by batteries. For long time use, you have to connect with an AC/DC adaptor. 2. Make sure sensor probe and meter are well connected. Don’t attempt to detach sensor probe from meter while in operation.
  • Page 6 “ATC” ICON INDICATION Probe type ntc 10K (Default) ntc 30K Plug in Temp. XX.X Temp. XX.X Manual Un-plugged “─ ─ ─” “─ ─ ─” temp. ATC icon CONDUCTIVITY/EC/TDS/SALT CALIBRATION PROCEDURE and MEASURING NOTE: Calibration is necessary before operation, the calibration procedures refer to as below “ a) to b)”: 2-1 Conductivity a) Required Equipment for calibration: 1) Cond./EC/TDS/SALT ELECTRODE...
  • Page 7 Conductivity electrode (K=10): Long press SET + CAL buttons to enter Cond. calibration mode. Short press MODE button to enter 12.8 mS calibration mode. Immerse the cond. probe (K=10) to standard 12880 us/cm buffer solution. Display will show the standard conductivity 12.8 mS value in the first layer. Long press SET button to finish calibration, screen will show “SA”.
  • Page 8 Factor= Actual TDS ÷ Actual Conductivity @ 25 ℃ Solution Factor Water 0.55~0.8 Drinking water 0.64 Absolute pure water 0.055 Distilled water 0.5 ~ 1.0 To enter the factor by long press SET+UNIT, then short press MODE to toggle Cont I FtdS. And go to select “tdS”and set the value.
  • Page 9: Electrode Maintenance

    CONDUCTIVITY STANDARD SOLUTION (1413 µS/cm at 25°C) ※Conductivity reading will be affected by the temperature. µS/cm µS/cm µS/cm ℃ ℉ ℃ ℉ ℃ ℉ 62.6 1198 93.2 1665 33.8 64.4 1224 1693 35.6 66.2 1251 96.8 1722 37.4 1277 98.6 1751 39.2 69.8...
  • Page 10: Sd Card Data Logging

    In most situations, hot water with a mild liquid detergent is an effective cleanser. Acetone easily cleans most organic matter, and hypochlorous solutions will remove algae,bacteria, or molds. To prevent cell damage, abrasives or sharp objects should not be used to clean an electrode. A cotton swab also works well for cleaning.
  • Page 11 5. When an SD card is used for the first time a folder is created on the card and named the model number. Under the MODEL number folder, the MODEL number and AUTO+YEAR folder will be automatically created. e.g.: /CDH-SD11/AUTO2017/….. 6. When data logging begins, a new folder named M(month)/D(date)/H(hour)/M(minute) is created on the SD card in the AUTO+YEAR folder.
  • Page 12 4. Data directory in SD card : /CDH-SD11/ MANUAL.csv Note: When manual data records full (199 points), logging will continue, but with new data overwriting old. If you want to keep the previous data, rename the file “MANUAL.csv” in /CDH-SD11/ MANUAL.csv is required.
  • Page 13: Troubleshooting

    BURN OUT NOTIFICATION If the sensor is not connected or over measurement range, the 「- - -」will be displayed on LCD and the data in the SD card is “0”. TROUBLE SHOOTING Q1: Wrong temperature?? A1: Refer to page 4 (TEMPERATURE PROBE TYPE), you must select the correct temperature sensor type or adjust temperature manually (Long press UNIT button then press UNIT to select “not”).
  • Page 14 Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification.