14.Using the IBASIC editor, enter the following program to clear the analyzer's ethernet
address. When entering the program, you must match upper/lower-case letters as shown.
10 OUTPUT 800; "SYST:COMM:LAN:EADD 'SimonSaysBeCareful'"
20 END
15.After the program has been entered, press
16.Using the IBASIC editor, enter the following program to install the analyzer's ethernet
address (recorded in
10 OUTPUT 716; "SYST:COMM:LAN:EADD '080009FA1234'"
20 END
17.After the program has been entered, press
18.If Option 100 is to be installed, press
Options Setup Install Options
recorded in
Step 4
keyboard connected to the analyzer's DIN KEYBOARD connector on the analyzer's rear
panel. If you don't have a keyboard, you can use the analyzer's arrow keys or the front
panel knob and the
analyzer's display. Press
19.Cycle power on the analyzer, and then press
Instrument Info
If you were not able load the analyzer's correction constants as instructed in
IMPORTANT
10, you must now perform the required adjustments that are described in
Chapter 3, Table 3-1, of the analyzer's Service Guide.
6
Step 5
or
Step
7). The actual ethernet address is not case sensitive.
. To enable the option, enter the keyword that was
or
Step
7. To enter the keyword, it is fastest and easiest to use a
key to select and enter characters from the top of the
Select Char
when done.
Enter
.Verify the analyzer's configuration is correct.
to execute.
Run
to execute.
Run