Understanding alarm indicators

The device will activate an alarm to alert you to conditions
that may require your attention or action. All the alarms on the
device are classified as low priority.
Each alarm will trigger:
a repeating audible tone
an illuminated yellow Alarm indicator on the control panel
an alarm message displayed in the Message field unless
noted differently below:
Alarm
Low battery
Critical low battery
Low oxygen purity
No breath detected
High temperature
Description
This alarm is triggered when the internal battery
has approximately a 5% charge remaining.
When the internal battery has a <5% charge
remaining, a countdown timer of the final 180
seconds of battery life will appear on the display
screen and an X will appear over the INT battery
bar.
This alarm is triggered when the delivered oxygen
concentration is lower than specified.
This alarm is triggered by a kinked cannula,
unattached cannula or shallow or mouth breathing.
This alarm is triggered when:
• the internal temperature of the device exceeds
149°F (65°C). The device will power off and the
audible alarm will cease. The yellow alarm
indicator and the alarm message will remain
until the device sufficiently cools.
• the temperature of the compressor exceeds
167°F (75°C). The device will continue running,
but the yellow alarm indicator and the alarm
message will remain until the device sufficiently
cools.