Endura AZ series integral and remote transmitter
Combustion oxygen monitor
Parameter
...Input/Output
...Digital I/O 1 (2)
Type
One point
Two point
Sequence
Calibration
Accuracy Check
Cal. On Error Limit
Warn On Error Limit
Table 5.12 Input/Output Menu (Continued)
IM/AZ20E–EN Rev. F
Comment / [Range]
Selects the type of automatic calibrations started by a
digital input – see Section 6.1, page 64.
Available only if the Input Function is a 'Start Autocal',
or 'Start/Stop Autocal' and the other I/O is not
Calibration Type.
Selects the type of calibration / check performed
when activated by a digital input.
Performs a calibration that updates the calibration
offset and factor automatically if successful.
Performs an accuracy check (the normal calibration
sequence) but does not update the calibration offset
and factor in use by the transmitter.
Performs an accuracy check but, if the error is greater
than the value set in the Cal. Error Limit field (see
page 41), the calibration offset and factor are
updated automatically.
Performs an accuracy check, but if the error is greater
than the value set in the Cal. Error Limit field (see
page 41) a calibration warning is generated.
5 Programming
Default
One point
Calibration
57