Input Module and Relay Card Installation

Before an Input Module can be inserted into the rear panel, a relay card must be installed.
Standard product configurations include at least one slot preconfigured with a relay card.
Use the procedure below and refer to Figure 2-6 as necessary to install the relay card:
1. Power off the Product with the main power switch.
2. Remove the four screws that secures the plastic frame to the rear panel.
3. Remove the plastic frame.
4. Slide the aluminum slot protector out of the Product.
5. Carefully align the rails of the relay card into the slot guides.
6. Slowly push the relay card into the Product until the card is fully seated.
Do not force the relay card into the slot. The card should easily
move when the rails of the relay card are properly aligned in the
slot guides.
7. Install the plastic frame and secure it to the rear panel with four screws.
2-8
Table 2-2. Instrument Setup Menu (cont.)
Description
Change the wait time or disable the screensaver.
Shows the date that the Product was last calibrated. See
the 1586A Calibration Manual for more information.
Change the Admin and User profile passwords.
Sets the Product to automatically turn on and resume
scanning and recording after a power loss.
Configure the LAN Ethernet or serial USB communication
settings.
Shows how many times the channel relays have been
scanned along with the name, serial number, and
calibration date of the Input Module.
Caution
Selections Available
Never, 15 Min, 30 Min, or
60 Min
--
See "Set Up Security" in this
chapter for more information.
See "Automatic Power Loss
Scan Resume" in Chapter 4
for more information.
See the 1586A Remote
Programmers Guide for more
information.
--