Trigger Input Termination - National Instruments MXI-2 Reference Manual

Hide thumbs Also See for MXI-2:
Table of Contents

Advertisement

Chapter 5
VXI-MXI-2 Configuration and Installation
Caution
Do not configure more than one MXIbus device to drive the MXIbus CLK10.
Having a second device driving MXIbus CLK10 could damage the device.
Drive CLK10 out MXIbus
Receive CLK10 from MXIbus
a. Receive CLK10 from MXIbus (Default)

Trigger Input Termination

MXI-2 Configuration Reference Manual
The VXI-MXI-2 can also drive or receive the MXIbus CLK10 signal.
Switch S7 controls whether the VXI-MXI-2 drives MXIbus CLK10 from
the VXIbus CLK10 or receives MXIbus CLK10. As shown in Figure 5-4c,
if W3 is configured to use the MXIbus CLK10 to generate the VXIbus
CLK10 signal, switch S7 must be configured to receive MXIbus CLK10.
This is shown again in Figure 5-6a. If you change the S7 setting to drive
CLK10 out the MXIbus, do not set the W3 jumper to receive the MXIbus
CLK10; instead, use the settings shown in Figure 5-4a or Figure 5-4b as
appropriate.
Figure 5-6 shows the configuration settings for receiving and driving
MXIbus CLK10, respectively.
Figure 5-6. Receiving or Driving MXIbus CLK10
You can use switch S2 to terminate the external trigger input SMB with
50 Ω. Figure 5-7a shows the default setting for a nonterminated trigger
input SMB. Use the setting of Figure 5-7b to terminate the trigger input
SMB. Switch S2 is above switches S3, S4, and S5, which have no effect on
this configuration.
Drive CLK10 out MXIbus
Receive CLK10 from MXIbus
b. Drive MXIbus CLK10 from VXIbus CLK10
5-10
ni.com