Specifications
Product Specifications
Standards Compliance
IM 701928-01E
Bandwidth*
Rise time*
Maximum continuous
input range
Maximum peak current
Output voltage rate*
Amplitude accuracy*
Noise*
Input impedance
Compatible interface
Temperature coefficient
for sensitivity*
Propagation delay
(Typical)
Maximum rated power
Rated supply voltage
Operating temperature
and humidity
Storage temperature
and humidity
Operating altitude
Effect of external
magnetic fields
Diameter of measurable
conductor
Guaranteed accuracy
period
Cable lengths
External dimensions
Weight
Accessories
* A ccuracy: 23 ±3°C, thirty minutes after turning ON the power. When used with a digital
oscilloscope having an input impedance of 1 MΩ ±1%.
Safety
EMC
Emissions
Immunity
PBC100 (701928): DC to 100 MHz (–3 dB)
PBC050 (701929): DC to 50 MHz (–3 dB)
(See the typical characteristics shown on page 11)
PBC100 (701928): 3.5 ns or less
PBC050 (701929): 7.0 ns or less
30 Arms (AC and DC components)
(See pages 11 and 12, Derating according to frequency)
50 Apeak, non-continuous
0.1 V/A
±1.0% rdg ±1 mV; 0 to 30 Arms
±2.0% rdg; 30 Arms to 50 Apeak
(DC, 45 to 66 Hz)
Equivalent to 2.5 mArms or less (with a 20 MHz bandwidth
measuring instrument)
(See typical characteristics on page 12)
YOKOGAWA probe interface
Within ±2% (input: 50 Hz, 30 Arms, within a range of 0 to
40°C)
701928: 13 ns
701929: 13 ns
PBC100 (701928): 5.5 VA (within maximum input range)
PBC050 (701929): 5.9 VA (within maximum input range)
±(12.3 ± 0.3) V, supplied from the YOKOGAWA probe interface
0 to 40°C 80% RH or less (no condensation)
–10 to 50°C, 80% RH or less (no condensation)
Up to 2000 m, indoors
Equivalent to a maximum of 5 mA (in a DC or 60 Hz, 400 A/m
magnetic field)
φ 5 mm
1 year (up to 10000 open and close operations)
Approx. 1.5 m (full length: approx. 1.755 m)
Sensor:
approx. 175 (W) x 18 (H) x 40 (D) mm
Probe interface: approx. 80 (W) x 30 (H) x 28 (D) mm
Approx. 190 g
User's manual, carrying case
EN61010
EN61326-1 Class B
EN55011 Class B, Group 1
EMC Regulatory Arrangement in Australia and New Zealand
EN 55011 Class B, Group 1
EN61326-1 Table 1 (Basic immunity requirement)
9