Using the Setup Assistant

6
To quickly set up the Astral device and
start ventilation for a new patient, use the
Setup Assistant.
1. Access Clinical mode.
2. From the main menu, press
3. Select Setup Assistant and press
Start.
4. A warning message will display. Select
Continue.
5. Select the patient type to automatically
configure alarm and setting ranges.
6. Select the circuit type to be connected.
7 . Select Start to run the Learn Circuit.
8. Follow the on-screen prompts to attach
and test the circuit.
9. Select your required ventilation mode.
The default settings for the mode are
displayed.
10. Review and adjust the settings as
required.
11. Press
For further information on adjusting
setting parameters and alarm settings, see
Adjusting patient settings and Adjusting
alarm settings in the Astral Clinical Guide.
.
to start ventilation.