Sunrise Medical Quickie Q700-UP M Instructions For Use Manual page 35

Powered wheelchair
Hide thumbs Also See for Quickie Q700-UP M:
Table of Contents

Advertisement

Available languages

Available languages

8.2.3 Drive wheel tyre repair
To remove the wheel/tyre:
Please see below and next page for photographs.
• Remove the colour ring.
• Use a 16 mm AF lug nut to loosen the 3-Socket studs, (Fig. 8.2).
• Jack the base/battery box up and support it using blocks,
(Fig. 8.3).
• Remove the 3 studs and pull the wheel away from the hub.
• Remove the valve cap and let the air out of the tyre by gently
pressing the valve stem with a small screw driver, (Fig. 8.4).
• There are 6 rim studs that must be loosened/tightened in the
number order shown, (Fig.8.5).
• Use a 5.0 mm Allen key to release/tighten the studs. (Fig. 8.5) .
• Lift the inner rim off the tyre wall (Fig. 8.6).
• Lift the tyre and tube off the outer rim, (Fig.8.7).
• Gently get a hold of the inner tube just behind the valve.
• Carefully feed the tube out of the tyre, (Fig.8.8).
• Ensure all parts are clean before reassembling, (Fig.8.9).
To refit
• Place the tube inside the tyre and rest the tyre on the outer
rim.
• Align the valve stem with the cut out in the rim.
• Face the valve stem outward.
• Place the inner rim over the tyre, tube and outer rim.
• Align the cut out to fit over the valve stem and match it up to
the cut out in the outer rim, (Fig.8.10).
• Make sure the stud mounts are in line on both rims.
• Tighten the studs in the sequence shown in Fig.8.5, taking care
not to pinch the tube.
• Slowly inflate to the pressures quoted in section 8.1.1 .
• Refit the wheel back onto the motor shaft and fix it safely with
the 3 studs to a torque of 34Nm.
For solid tyres the sequence is the same, just discount the
references to the inner tube and valve.
Fig. 8.2
Q700-UP M/F Rev.A
Fig. 8.3
Fig. 8.4
Fig. 8.5
Fig. 8.6