HP 53132A Assembly And Service Manual page 64

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 1 Performance Tests
HP 53131A/132A Performance Test Record (Page 2 of 6)
HP 53131A/132A Performance Test Record
(Page 2 of 6)
Test
Number
Test Description
1
Time Interval
(HP 53131A/132A
Only):
For HP 53131A
For HP 53132A
2
Trigger Level:
Channel 1
Channel 2
(HP 53131A/132A Only)
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
Standard (Front Panel Terminals)
100 kHz–100 MHz, Channel 1
100 kHz–100 MHz, Channel 2
100 MHz–200 MHz, Channel 2
100 MHz–200 MHz, Channel 1
200 MHz–225 MHz, Channel 1
200 MHz–225 MHz, Channel 2
Option 060 Rear Terminals
100 kHz–225 MHz, Channel 1
100 kHz–225 MHz, Channel 2
1-42
Complete Performance Tests
Minimum
0.1460 µs
0.1475 µs
−15 mV
−15 mV
Actual Reading
1. __________
2. __________
3. __________
4. __________
5. __________
6. __________
Actual Reading
1. __________
2. __________
Actual Reading
Maximum
0.1540 µs
1. __________
0.1525 µs
2. __________
1. __________
+15 mV
2. __________
+15 mV
Specification
≤ −21 dBm (20 mVrms)
≤ −21 dBm (20 mVrms)
≤ −17.5 dBm (30 mVrms)
≤ −17.5 dBm (30 mVrms)
≤ −15.0 dBm (40 mVrms)
≤ −15.0 dBm (40 mVrms)
Specification
≤ −9.5 dBm (75 mVrms)
≤ −9.5 dBm (75 mVrms)
Assembly-Level Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53181a53131a

Table of Contents