To Calibrate The Standard Timebase - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 2 Service
The HP 53131A/132A Calibration Procedures

To Calibrate the Standard Timebase

1
Connect the output of a 10 MHz house standard to Channel 1 of
the Counter as shown in Figure 2-5.
HP 53131A/132A
Counter
Channel 1
Figure 2-5. Standard Timebase Calibration Setup
2
Cycle the POWER key to preset the Counter.
Allow the counter to warm up for approximately 30 minutes.
3
With an insulated tuning tool, turn the OSC Adjust potentiometer
(located on the rear chassis of the Counter) until the frequency
reading in the Counter display is 10.00000000 MHz ± 5 Hz.
The Counter will be calibrated to the house standard.
If you are not able to adjust the frequency correctly, refer to the
troubleshooting section in this chapter.
This completes the Standard Timebase Calibration procedure.