Cal Volts Input - Fluke 724 Calibration Manual

Hide thumbs Also See for 724:
Table of Contents

Advertisement

724/725
Calibration Manual

Cal Volts Input

Connect the UUT as shown in Figure 13.
Warning
30 V maximum to
SOURCE / MEASURE
mA +
3W
mA -
4W
From the Calibration Menu, type the cal step for Cal Volts Input. The PC displays:
Set the Fluke 5520A to output 0.0000 V, then press the space bar. After a short while,
the following calibration constant and prompt are displayed on the PC:
Set the Fluke 5520A to output 10.0000 V, and press the space bar. After a short while,
the following calibration constants and prompt are displayed on the PC:
Press the space bar to return to the Calibration Menu.
34
30V MAX ALL TERMINALS
MEASURE
V
V
TC
mA
Hz
RTD
LOOP
TC
COM
COM
UUT
Figure 13. Volts Input Calibration Connections
Enter 0 Volts
- press space bar to continue
Offset =
-40
Enter 10.00 Volts
- press space bar to continue
diff = (Counts - Offset)
3032676
=
3032636
Volts per count =
0.000003
- press space bar to continue
-
-40
NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
HI
1000V
20V
1V PK
RMS
RMS
MAX
MAX
MAX
LO
20V
RMS
MAX
SHELLS
20A
GUARD
NOT
GROUNDED
20V PK MAX
TC
20V PK MAX
5520A
SCOPE
OUT
150V
PK
MAX
TRIG
20V
PK
MAX
aal04f.eps