Thank you for purchasing the BUDDI Read me Please take a few minutes to read this user manual. first If you have any questions at any time, please contact NanoEnTek Inc. ,customer support. Tel: +82-2-6220-7942 Fax: +82-2-6220-7999 E-mail: ivdst@nanoentek.com 20160901 BUDD .indd 2...
Na- noEnTek Inc., not regularly maintained, used with equipment not ap- proved by NanoEnTek Inc. or used for purposes for which it was not designed. This warranty does not apply to damages incurred during the initial shipment of the instrument and associated auxiliary materials.
Introduction The principle of the BUDDI Principle BUDDI is an image analyzer that evaluates the line signal intensities • of analyte and control line on the test device. • BUDDI measures the intensity of analyte and control line on the test device, and then interprets the results using scoring algorithm.
Introduction BUDDI™ is shipped with the following components. Package List Please check all components listed below when an instrument was de- livered. If any item is missing or damaged, contact your local distributor or e- mail ivdst@nanoentek.com. BUDDI Adaptor / Power cord...
• Leave adequate distance between the back of the instrument and the wall to allow the operation of the ‘on’ and ‘off’ power switch. Do not insert metallic objects into the BUDDI as this could result • in an electrical shock, personal injury and equipment damage. Insert only qualified and authorized devices supplied by NanoEnTek Inc.
Introduction The following symbols are found on the BUDDI™ and in the manual General and package inserts. Study the meanings of the symbols and always Safety use the equipment in a safe manner. Warnings • Before using the equipment, read this manual and be certain you thoroughly understand it’s contents.
Introduction BUDDI Name Function description Device Slot Location to insert or remove the BUDDI device Power On/Off Power Switch (Press and hold the power button for 3 seconds to turn the power ON.) Screen Show the operating status and results USB Port...
Page 12
Introduction Test screen Buttons & functions Designation Purpose Preview the device Patient ID & user ID input field Start a test. Cancel a test and move to main screen. 20160901 BUDD .indd 12 2016-09-05 2:22:18...
Introduction Setting screen Buttons & functions Designation Purpose Set the time and date at which the BUDDI™ is located. Calibrate the BUDDI ’s alignment, test line and control line, to enhance detection performance Set a e-mail environment setting. Set your e-mail server address, port and your mail account.
Setup 1. Carefully take out the BUDDI™ and other accessories from the box. BUDDI 2. Connect the power cord and adapter, and connect the adapter to the setup rear side of the BUDDI™. procedure 3. Plug the power cord into the power outlet.
Setup 12. If you want to connect wireless internet, press the Wifi button on BUDDI the setting screen. setup procedure 13. Selects the appropriate SSID. If the SSID does not appear, push the Search button. Then the SSID list will be displayed. Input password and encryption options, press the connect button.
NOTE: The test device, added a patient sample, should be tested within 15 minutes. Do not interpret a test result after 15 minutes. 4. Insert the device into the device slot of BUDDI™. 5. Press the TEST button on the Main screen to move to the TEST screen.
Operation BUDDI 7. When the test is done, the results are displayed on the screen. operation procedure 8. If you want to save the result to the connected USB drive, press the USB button. 9. If you want to send the result to the LIS server, press the LIS button.
Page 22
8. You will see “Pass”, if the system is properly operating. 9. If you want to save the calibration result list, press the USB button. 10. Press the CLOSE button and move to the Main screen. 11. Remove the calibration device from BUDDI™. 20160901 BUDD .indd 22...
For safety reasons, disconnect the power supply from the main • outlet and remove the power adapter equipment from the BUDDI • If you want to store for a lengthy period, keep the BUDDI covered in a dry place. • Do not expose the equipment to direct sunlight.
Maintenance Trouble- If the BUDDI malfunctions, please check the following before you contact Customer Support. shooting Error Possible Type Error message Possible Solution code Cause The device Re - i n s e r t t h e D ev i c e...
The image is not Cannot detect Check the control line on EM-07 detected. Please the control line the device try again Cannot find BUDDI.exe file is “BUDDI.exe” Check the “BUDDI.exe” EM-08 not exist file in USB file in the USB drive drive Update...
Procedure Maintenance General • Electrical rating: Voltage: AC100 ~ 240 V, Current: 1.7 A, Frequency: 50 ~ 60 Hz specification • Electric input: Voltage: DC 12 V, Current: 5 A -normal • Dimensions: 166mm x 202 mm x 205 mm (W x L x H) environmental •...
Every effort has been made to avoid errors in test and diagrams. How- ever, NanoEnTek Inc. assumes no responsibility for any unintentional errors which may appear in this publication. Information and specifica- tions contained in this document are subject to change at any time without prior notice.