Download Print this page

Teledyne 402 Instruction Manual

Analytical instruments

Advertisement

Quick Links

Equipment needed:
Document needed:
Gases needed:
Zero gas: Zero grade (representative of background gas)
Fuel: (40% H2 / 60% N2) or 100% H2
Air: Zero grade air
Span gas: (70-90% of range) ppm CH4 balance back ground)
Procedure:
1. Connect power cord to AC inlet
2. Position ON / OFF switch to ON
3. Connect zero grade air to analyzer, set the input pressure to 30 PSIG, adjust the air regulator inside the
sample module for 15 psig or as indicated in the instruction manual
4. Connect Zero gas to analyzer, set the pressure to 30 PSIG or same as sample pressure, adjust sample
bypass valve so the sample pressure gauge inside the sample module indicates the pressure in the
instruction manual, and sample bypass flow meter reads 0.5 to 2 SCFH
5. Connect fuel (40 / 60 mix or 100 H2 per application); set pressure to 30 PSIG
6. Check 0-1 VDC, 4-20 MA output, and alarm(s), by adjusting the zero and span pot to get the analyzer
reading from 0-100% of full scale
7. Let the analyzer run for 8 hours or overnight
8. After the sample cell temperature is stabilized, position range switch to Ignite; hold for 6 seconds, then
release to HI range
9. Wait until the analyzer reading is stabilized, introduce span gas, adjust fuel regulator in increments of
0.5 PSIG at a time, and watch for the analyzer reading to increase; continue to adjust fuel regulator until
the analyzer reading is increasing slower or the reading starts to decrease
10. Wait until the analyzer reading is stabilized, adjust span pot until the analyzer reads the concentration
in the cylinder
11. Introduce zero gas, wait until the reading is stabilized, adjust zero pot until analyzer reading is 0 ppm
12. Repeat steps 9 and 10 until no further adjustment is required
13. Connect sample to the analyzer, check the sample pressure and flow rate; now the analyzer is ready to
go on line
Model 402, 402R, 402REU: Start-up Procedure
Digital multimeter (DMM)
Power cord
Instruction manual
TELEDYNE ANALYTICAL INSTRUMENTS
www.teledyne-ai.com

Advertisement

loading

Summary of Contents for Teledyne 402

  • Page 1 TELEDYNE ANALYTICAL INSTRUMENTS www.teledyne-ai.com Model 402, 402R, 402REU: Start-up Procedure Equipment needed: Digital multimeter (DMM) Power cord Document needed: Instruction manual Gases needed: Zero gas: Zero grade (representative of background gas) Fuel: (40% H2 / 60% N2) or 100% H2...
  • Page 2 TELEDYNE ANALYTICAL INSTRUMENTS www.teledyne-ai.com NOTE: The flow restructures are: Blanket air = 600 sccm Sample = 30 sccm Fuel = 200 sccm (Standard 40 / 60 mix) Fuel = 30 sccm (100% H2) When the background is hydrogen, the fuel port is connected to zero grade nitrogen...

This manual is also suitable for:

402r402reu