Download Print this page
Omron HEM-7130 Instruction Manual

Omron HEM-7130 Instruction Manual

Automatic blood pressure monitor

Advertisement

Quick Links

1. Overview
Main unit:
A
Automatic Blood
D
F
Pressure Monitor
E
D D
Model HEM-7130
Instruction Manual
A. Display
B. Memory button
5342276-9A
C. START/STOP button
D. Date/Time setting (
) button
E. Cuff wrapping guide lamp
Introduction
Thank you for purchasing the OMRON HEM-7130 Automatic Blood Pressure
Arm cuff:
Monitor.
The OMRON HEM-7130 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
"IntelliSense" technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
Display:
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
N
O
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
P
For specific information about your own blood pressure, CONSULT YOUR
Q
DOCTOR.
R
S
Important Safety Information
T
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
U
or arteriosclerosis.
Please read this section carefully before using the unit.
N. Average value symbol
O. Memory symbol
Warning: Indicates a potentially hazardous situation which, if not avoided,
P. Irregular heartbeat symbol
could result in death or serious injury.
Q. Systolic blood pressure
R. Diastolic blood pressure
(General Usage)
S. Low battery symbol
• Do not use the device on the injured arm or the arm under medical treatment.
T. Cuff wrapping guide
• Do not wrap the arm cuff on the arm while being on an intravenous drip or
blood transfusion.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
2. Preparation
the power and unplug the power cord immediately.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
2.1 Installing/Replacing the Batteries
multiple-tap.
• Never plug in or unplug the power cord from the electric outlet with wet
1.
Remove the battery cover.
hands.
2.
Caution: Indicates a potentially hazardous situation which, if not avoided,
Insert four "AA" batteries
may result in minor or moderate injury to the user or patient or
as indicated in the battery
damage to the equipment or other property.
compartment and then
replace the battery cover.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
Notes:
• People with severe blood flow problems, or blood disorders, should consult a
• If the low battery symbol (
doctor before using the unit, as cuff inflation can cause internal bleeding.
monitor off then replace all batteries at the same time.
• If there are any abnormalities during the measurement, remove the arm cuff.
• The measurement values continue to be stored in memory even after
• Do not use this device on infants or persons who cannot express their
the batteries are replaced.
intentions.
• The supplied batteries may have a shorter life.
• Do not inflate the arm cuff more than necessary.
Disposal of used batteries should be carried out in accordance with the
• Do not use the unit for any purpose other than measuring blood pressure.
national regulations for the disposal of batteries.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
2.2 Setting the Date and Time
• Do not use a mobile phone or other devices that emit electromagnetic fields,
1.
near the unit. This may result in incorrect operation of the unit.
Press the
button.
• Do not disassemble the unit or arm cuff.
2.
Set the monitor to the correct date and time before
(Battery Usage)
taking a measurement for the first time.
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only four "AA" alkaline or manganese batteries with this unit. Do not use
other types of batteries. Do not use new and used batteries together.
• Remove the batteries if the unit will not be used for 3 months or more.
(AC Adapter (optional) Usage)
• Fully insert the power plug.
• When disconnecting the power plug, do not pull the power cord. Be sure to
hold the power plug.
• When handling the power cord, observe the following:
Do not damage.
Do not break it.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not bundle during use.
Do not pinch.
Do not place under heavy objects.
3.
Press the START/STOP button to store the setting.
• Wipe the dust off from the power plug.
• Disconnect the power plug if the product will not be used for a long period of
Notes:
time.
• If the batteries have been replaced, the date and time setting will
• Disconnect the power plug before starting maintenance.
need to be reset.
• Use only the original AC adapter designed for this unit. Use of unsupported
• If the date and time are not set, "-:--" appears during or after
adapters may damage and/or may be hazardous to the unit.
measurement.
General Precautions
3. Using the Unit
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• To unplug the air plug, pull on the air plug at the connection with the main
3.1 Applying the Arm Cuff
unit, not the tube itself.
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
Remove tight-fitting clothing or tight rolled up sleeve from your
• Do not inflate the arm cuff when it is not wrapped around your arm.
upper arm. Do not place the arm cuff over thick clothes.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the Technical Data Section.
1.
Insert the air plug into the air jack
• Read and follow the "Correct Disposal of This Product" in the Technical Data
securely.
Section when disposing of the device and any used accessories or optional
parts.
2.
Put your arm through the cuff
loop.
G
H
I
B
The bottom edge of the arm
cuff should be 1 to 2 cm
C
above the elbow. The air
F. Up/Down (
) buttons
tube is centred on the
G. Battery compartment
middle of your inner arm.
H. AC adapter jack (for optional
AC adapter)
I. Air Jack
3.
Close the fabric fastener
FIRMLY.
J
J. Arm Cuff (Arm circumference
22 - 42 cm)
K. Air plug
L. Air tube
K
L
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on
V
the air tube.
W
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
X
measurement. If the values between both arms differ substantially,
please check with your doctor which arm to use for your
Y
measurement.
Z
3.2 How to Sit Correctly
To take a measurement, you need to be
U. Heartbeat symbol
• Flashes during measurement.
relaxed and comfortably seated, under
V. Date/Time display
comfortable room temperature. No
W. Movement error symbol
bathing, drinking alcohol or caffeine,
X. Blood pressure level indicator
smoking, exercising or eating 30 minutes
Y. Deflation symbol
before taking a measurement.
Z. Pulse display and Memory
number
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The arm cuff should be at the same level as
your heart.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button to release
the air in the arm cuff.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
START
INFLATE
DEFLATE
) appears on the display, turn the
Cuff Wrapping Guide Lamp
If your systolic pressure is more than 210 mmHg
1)
Push
or
buttons to
After the arm cuff starts to inflate, press and hold the START/STOP
change.
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
2)
Push the
button to confirm.
Next setting appears.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Note: Wait 2-3 minutes before taking another blood pressure
Year
measurement. Waiting between readings allows the arteries to
Month
Day
Hour
Minute
return to the condition prior to taking the blood pressure
measurement.
Always consult your doctor. Self-diagnosis of measurement
results and self-treatment are dangerous.
Cuff Wrapping Guide Lamp:
The cuff wrapping guide lamp
lights in green, the cuff is correctly
wrapped tightly enough on the arm and the reading is accurate and
reliable. If it does not light, the reading is not reliable due to the
incorrect wrapping of the cuff. Please wrap the cuff again, taking care
to wrap it correctly and take the measurement again.
2.
Remove the arm cuff.
3.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Important:
• Recent research suggests that the following
4.1 The Icons and Error Messages
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
1 to 2 cm
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement.
The irregular heartbeat algorithm automatically
determines if the measurement is usable or
needs to be repeated. If the measurement
results are affected by irregular heartbeats but the result is valid, the
result is shown together with the irregular heartbeat symbol (
). If
the irregular heartbeats cause the measurement to be invalid, no
result is shown. If the irregular heartbeat symbol (
) is shown after
you have taken a measurement, repeat the measurement. If the
irregular heartbeat symbol (
) is shown frequently, please make
your doctor aware of it.
• If you move during measurement, the movement
error symbol (
) will appear on the display.
Keep still and repeat the measurement.
Using the Guest Mode
The monitor stores measurement values for one user in the memory.
The guest mode can be used to take a single measurement for
another user. No measurement values are stored in the memory when
the guest mode is selected.
1.
Press and hold the START/STOP
button for more than 3 seconds.
The date/time display will disappear.
2.
Release the START/STOP button when the date/time
display turns off.
The arm cuff will start to inflate automatically.
3.4 Using the Memory Function
The monitor automatically stores the result up to 60 sets. It can also
calculate an average reading based on the measurements from the
last three readings taken within 10 minutes. If there are only two
readings in memory for that period, the average will be based on two
readings. If there is one reading in memory for that period, the average
will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest reading.
• When viewing the reading taken without setting the date and time,
"-:--" is displayed instead of the date and time.

4.2 Troubleshooting

To View the Readings Stored in Memory
COMPLETE
1.
Press the
button.
The Memory number appears for a second before the pulse rate is
displayed. The newest set is numbered "1".
Alternating date/time display
Note: The cuff wrapping guide result appears
on the display with the measurement
values. The cuff wrapping guide lamp will
not light.
: The arm cuff was correctly wrapped tightly enough.
: The arm cuff was loose or not wrapped correctly.
2.
Press the
or
button to view the readings stored in
memory.
: To the older readings
: To the more recent readings
To View the Average Value
1.
Press and hold the
button for
more than 3 seconds.
Notes:
• If the previous reading was taken without setting
the date and time, the average value is not
calculated.
• If there are no measurements results stored in
the memory, the screen to the right is displayed.
To Delete All the Values Stored in Memory
When the memory symbol (
) appears, first press the Memory
button. Then while holding it down, press the START/STOP
button simultaneously for more than 3 seconds.
Note: You cannot partially delete the stored readings.
4.3 Maintenance
4. Troubleshooting and Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme
Error Display
Cause
Remedy
temperatures, humidity, moisture or direct sunlight.
Remove the arm cuff. Wait
• Do not fold the cuff or tubing tightly.
2 - 3 minutes and then take
• Do not inflate the arm cuff over 299 mmHg.
Irregular heartbeats are
another measurement.
• Do not disassemble the unit.
detected.
Repeat the steps in section
• Do not subject the unit to strong shocks or vibrations
3.3. If this error continues to
(for example, dropping the unit on the floor).
appear, contact your doctor.
• Do not use volatile liquids to clean the main unit.
Movement during
Carefully read and repeat
• Do not wash the arm cuff or immerse it in water.
measurement.
the steps in section 3.3.
• Do not use petrol, thinners or similar solvents to clean
You should replace them
the arm cuff.
with new ones ahead of
The batteries are low.
• Do not carry out repairs of any kind yourself. If a defect
time.
Refer to section 2.1.
occurs, consult your local OMRON representative.
You should replace them
The batteries are
with new ones at once.
exhausted.
Refer to section 2.1.
Insert the plug securely.
Air plug disconnected.
Refer to section 3.1.
Arm cuff not applied
Apply the arm cuff correctly.
correctly.
Refer to section 3.1.
Replace the cuff with the
Air is leaking from the arm
new one.
cuff.
Refer to Chapter 5.
Repeat measurement.
Remain still and do not talk
• The unit should be cleaned with a soft, dry cloth.
during measurement.
• Use a soft, moistened cloth and neutral soap to clean
Refer to section 3.3.
Movement during
the arm cuff.
measurement and the arm
If "E2" appears repeatedly,
cuff has not been inflated
inflate the cuff manually
sufficiently.
Note: Read and follow the "Correct Disposal of This Product" in the
until it is 30 to 40 mmHg
Technical Data Section when disposing of the device and any
above your previous
used accessories or optional parts.
measurement result.
Refer to section 3.3.
Calibration and Service
The arm cuff was inflated
Do not inflate the cuff above
• The accuracy of this blood pressure monitor has been
above 299 mmHg
299 mmHg.
carefully tested and is designed for a long service life.
when inflating the cuff
Refer to section 3.3.
manually.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
Repeat measurement.
accuracy. Please consult your local OMRON
Movement during
Remain still and do not talk
measurement.
during measurement.
representative.
Refer to section 3.3.
Remove any clothing
Clothing is interfering with
interfering with the arm cuff.
the arm cuff.
4.4 Storage
Refer to section 3.1.
Contact your local OMRON
1.
Unplug the air plug from the air jack.
Device error.
representative .
2.
Gently fold the air tube into the arm cuff.
Notes:
• Do not bend the air tube excessively.
Problem
Cause
Remedy
• Do not store the unit in the following situations:
Apply the arm cuff
- If the unit is wet.
Arm cuff not applied
correctly.
- Locations exposed to extreme temperatures, humidity, direct
correctly.
Refer to section 3.1.
sunlight, dust or corrosive vapours.
- Locations exposed to vibrations, shocks or where it will be at an
Remain still and do not
angle.
The reading is
talk during
Movement or talking
extremely low (or
measurement.
during measurement.
high).
Refer to section 3.3.
Remove any clothing
Clothing is interfering
interfering with the arm
with the arm cuff.
cuff.
Refer to section 3.1.
Make sure that the air
5. Optional Parts
The air tube is not
tube is connected
securely connected
securely.
Wide Range Arm Cuff
into the main unit.
Arm cuff pressure
Refer to section 3.1.
Arm circumference
does not rise.
22 - 42 cm
Replace the arm cuff
Air is leaking from the
with a new one.
arm cuff.
Refer to Chapter 5.
The cuff wrapping
Apply the cuff correctly
guide lamp does not
so that it is firmly
The arm cuff is loose.
light.
wrapped around the arm.
Arm cuff deflates too
Refer to section 3.1.
Note: Please check with your local OMRON representatives for the
soon.
appropriate optional part models.
Inflate the cuff so that it
is 30 to 40 mmHg above
Cannot measure or
The arm cuff has not
Using the Optional AC Adapter
your previous
readings are too low
been inflated
measurement result.
or too high.
sufficiently.
1.
Insert the AC adapter plug
Refer to section 3.3.
into the AC adapter jack on
Replace the batteries
the rear side of the main
The batteries are
with new ones.
empty.
unit.
Nothing happens
Refer to section 2.1.
when you press the
2.
Plug the AC adapter into an
The batteries have
Insert the batteries with
buttons.
been inserted
the correct (+/-) polarity.
electrical outlet.
incorrectly.
Refer to section 2.1.
To disconnect the AC adapter, unplug the AC adapter from the
• Press the START/STOP button and repeat
electrical outlet first and then remove the AC adapter plug from the
measurement.
main unit.
• If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
6. Technical Data
Product Description
Automatic Blood Pressure Monitor
Model
HEM-7130
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/ min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Memory
60 Measurements with date and time
Rating
DC6V 4W
Power Source
4 "AA" batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Battery Life
Approx. 1000 measurements
(using new alkaline batteries)
Applied Part
= Type BF
Protection Against
Internally powered ME equipment (When using only
Electric Shock
the batteries)
= Class II ME equipment (Optional AC adapter)
Operating
+10 to +40°C / 30 to 85% RH
temperature/ Humidity
Storage temperature/
-20 to +60°C / 10 to 95% RH / 700-1060 hPa
Humidity/ Air pressure
Console Weight
Approx. 280g without batteries
Cuff Weight
Approx. 170g
Outer Dimensions
Approx. 107 (w) mm × 79 (h) mm × 141(l) mm
Cuff Dimensions
Approx. 151 mm × 563 mm
(Cuff: arm circumference 22 to 42 cm)
Cuff/ Tube Material
Nylon, polyester, polyvinyl chloride
Package Contents
Main unit, arm cuff, instruction manual, battery set
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
AC Adapter S
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS
www.omron-healthcare.com
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omronhealthcare-ap.com
Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM
CO., LTD.
Binh Duong Province, VIETNAM
Made in Vietnam

Advertisement

loading