ADC e-sphyg 2 User Manual page 16

Automatic sphygmomanometer
Hide thumbs Also See for e-sphyg 2:
Table of Contents

Advertisement

Apply the cuff and stethoscope to the patient's arm in the same
manner as the usual auscultatory method.
Press START/STOP button.
The cuff is inflated automatically after zero settings.
Pressurization stops when pressure reaches the preset pressure value and
pressure starts to descend.
Deflation rate is displayed at the lower right corner of the segment LCD.
NOTE: Press START/STOP button whenever the measurement
should be interrupted. The device will rapidly exhaust air from the cuff.
If cuff inflation was not sufficient, press and hold down
START/STOP button once deflation has started and the cuff
is pressurized again. Pressurization will continue while the
button is pressed. Pressure does not exceed 290 mmHg.
Press START/STOP button when blood pressure is taken and air
is rapidly exhausted from the cuff.
Air is also rapidly exhausted from the cuff when pressure
descends to 30 mmHg.
Turn off the device by pressing " " side of MAIN POWER SWITCH.
The device will be in the energy saver mode after approximately
3 minutes if it is not turned off. See page 10.
Make sure your patient is relaxed, has taken at least 5 minutes of rest before
measurement, and refrains from talking or moving during measurement.
Confirm that the cuff is kept at the height of patient's heart during measurement.
Do not execute repeated measurements for congestion of blood could result
in false measurement. Allow the patient's arm to rest for at least 5 minutes.
16