Electromagnetic Immunity

The Fluke 43, including standard accessories, conforms with the EEC directive
89/336 for EMC immunity, as defined by IEC1000-4-3, with the addition of the
following tables.
Disturbance with test leads TL24 or Current Clamp 80i-500s
Volts / amps / hertz
Resistance, Capacitance
Power
Harmonics
No visible disturbance
Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 1 GHz
(-): no visible disturbance
Disturbance with test leads TL24 in scope mode
V AC+DC (True RMS)
Disturbance less than 1 %
of full scale
Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 200 MHz
Frequency: 200 MHz - 1 GHz
(-): no visible disturbance
Disturbance less than 10 % of
full scale
Frequency: 10 kHz - 27 MHz
Frequency: 27 MHz - 200 MHz
Frequency: 200 MHz - 1 GHz
(-): no visible disturbance
Ranges not specified in Tables 2 and 3 may have a disturbance of more than
10 % of full scale.
Specifications
Electromagnetic Immunity
Table 1
E = 3 V/m
(-)
(-)
Table 2
E = 3 V/m
2 V/div - 500 V/div
500 mV/div - 500 V/div
(-)
Table 3
E = 3 V/m
1 V/div
200 mV/div
(-)
3
E = 10 V/m
(-)
(-)
E = 10 V/m
10 V/div - 500 V/div
2 V/div - 500 V/div
5 mV/div - 500 V/div
E = 10 V/m
5 V/div
1 V/div
(-)
41