*ESR?
Bit
0
1
2
3
4
5
6
7
*IDN?
*OPC
*OPC?
*OPT?
*RST
1.800.561.8187
Return and clear the IEEE-488.2 Standard Event Status Register.
The response is a decimal number that ranges from 0 to 255. The definition of
the bits are shown in Table 5.
Name
OPC
RQC
QYE
DDE
EXE
CME
URQ
PON
Return product information. The response contains the manufacturer, model
number, serial number, and firmware version numbers.
Operation complete. Since all commands are sequential, this command is not
necessary for synchronization.
Return the product configuration.
The response is 1 if the Product is equipped with the Input Module (7109A-P or
6109A-P only). It is 0 otherwise.
Reset operational settings.
Stop temperature control
Set the setpoint to 25 °C
Set Control Sensor to Internal
Set Ramp Enable to OFF
Set Stability Beep to ON
Set Stability Window to 0.05 °C
Stop program execution
www.
Table 5. *ESR Bits
Definition
Operation complete
Request control (not used)
Query error flag (not used)
Device-specific error flag
Execution error flag
Command error flag
User request (not used)
Power on