Type
PTGC + Vortex
Strip gas
Nitrogen
Table 3-6
Influencing variables
Effects of ambient temperature
Table 3-7
Mounting Clearance
Mounting clearance
Table 3-8
Dimensions
Weight
Overall Dimensions
(including mounting hardware)
Maxum II Reference Manual
Manual, 7/2017, 2000596-001
Quality
Pressure
Instrument air
400 to 600 kPa (58 to 87
psig)
5.0 (≥ 99,999%)
400 to 600 kPa (58 to 87
psig)
Negligable with electronic pressure control. Different effects with mechanical pressure
control (depending on application)
● Clearance on left: 460 mm (18") from walls and other devices
● Clearance on right: 460 mm (18") in all cases
● Clearance at front: 654 mm (26") in all cases
● Center-to-center: 1,120 mm (44") in all cases
77 kg (170 lbs.) application dependent
Height: 1053.6 mm (41.5")
Width: 752 mm (26.6")
Depth: 417.4 mm (16.4")
Specifications
Typical Consumption
Approx. 575 l/min
6 to 8 l/h