Accu-Chek ADVANTAGE User Manual page 56

Hide thumbs Also See for ADVANTAGE:
Table of Contents

Advertisement

IF...
the following screen is displayed.
REASON
The meter has failed its internal system check.
ACTION
Remove the Chek
Strip. Turn the meter off and
repeat the glucose control check. See Glucose
Control Testing
(page 27).
appears, testing cannot take place. Call Accu-
Chek
SM
Customer Care at 1-800-858-8072.
108
If this error display still
IF...
the following screen is displayed.
REASON
The test strip may be damaged or the test was not
performed correctly.
ACTION
If this display appears before blood is placed on the
strip, remove the test strip and reinsert. If this
display appears after blood is applied to the strip,
there is an incorrect amount of blood on the strip.
Repeat the test with a new test strip.
109