Errors and warning messages

19. Errors and warning messages
Memory full
Reference memory full
You will also find an error number for the following messages in the Calibration/
Status menu to provide support for diagnostics:
Tolerance
Error 01
Please call Service
Error 02
Error 03
Error 04
Error 05
Please observe the instructions on cleaning standards in the section on
Calibration.
50
Transfer the content of memory to a PC and then
delete the contents of memory.
A maximum of 50 references can be saved. It may
be necessary to delete old references.
Generally occurs only with major changes in
climatic or weather conditions. The deviation was
successfullycompensated for by calibration and
correct measurements are still possible. You should
still recalibrate the device as soon as it is operating
in normal climatic conditions again. However if a
change in climate cannot be considered as the
cause of the problem, you should check whether the
standard is clean.
Autodiagnosis has determined an impermissible
Service (invalid) deviation in the measurement signal
that cannot be remedied by recalibrating.
Generally occurs when there is a significant amount
of dirt or dust on the standard or optics. First try to
clean the standard. You should only have the optics
cleaned by our Customer Service department, for
example as part of a yearly recertification.
Defect in the electronics or operating error. First
check whether the standard is clean and whether
the device is properly snapped into the holder.
Defect in the lamp or electronics.
Defect in the electronics.