Appendix A: Technical Specifications
Category
Maternal Pulse Oximetry Mode (Nellcor)
Technique:
Sensor Type and Accuracy
®
OxiMax
Sensor Models
2
MAX-A
,
DS-100A
Saturation Range:
Pulse Rate Range:
Accuracy:
Saturation (SpO
%)
2
2
Adults
3
Low Perfusion
Pulse Rate (bpm)
Adults
4
Wavelengths
:
Red
Infrared
Maximum Optical Output Power:
Response Time:
Alarms (audible and visual):
Audio
Visual
Limits
Technical
1) Accuracy specifications are based on controlled hypoxia studies with healthy, non-smoking adult volunteers over the specified saturation SpO
range. Pulse oximeter SpO
readings were compared to SaO
2
as ± "X" digits. This variation equals ± one standard deviation (± 1 SD), which encompasses 68%of the population. Oxygen saturation accuracy can be
affected by certain environmental and patient physiological conditions, as discussed in the operator's manual for the monitor. Use Nellcor sensors only
with 250cx Series Monitors containing Nellcor oximetry. Consult individual manufacturers for accuracy specifications and compatibility information
of particular instruments and Nellcor sensor models. The volunteer population was composed of healthy men and women recruited from the local
population. The ages ranged from 18 to 50 years old, with variations of skin pigmentations.
2) Adult specifications are shown for OxiMax
®
3) Applicability: OxiMax
MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
4) Information of wavelength range can be especially useful to clinicians performing photodynamic therapy.
NOTE: Because pulse oximeter equipment measurements are statistically distributed, only about two-thirds of pulse oximeter equipment measurements
can be expected to fall within ±
NOTE: Use of a functional SpO
2
This device is covered under one or more of the following Patents: US Patent No. 4,802,486; 4,869,254; 4,928,692; 4,934,372; 4,960,126; 5,078,136;
5,485,847; 5,743,263; 5,865,736; 6,035,223; 6,298,252; 6,463,310; 6,591,123; 6,675,031; 6,708,049; 6,801,797; Re. 35,122; and foreign equivalents.
140
Operating Modes Specifications (Continued)
1
:
values of drawn blood samples measured by hemoximetry. All accuracies are expressed
2
®
MAX-A and MAX-N sensors with the N-600. Saturation accuracy will vary by sensor type.
A
rms of the value measured by a CO-Oximeter.
simulator to assess the accuracy of the Corometrics
2036947-001
Spectrophotometry and plethysmography.
SpO
Range:
2
70%–100%:
± 2 digits
± 3 digits
1–100%
30–250 bpm
70%-100% ± 2 digits
70%-100% ± 2 digits
0%-69% unspecified
20 to 250 bpm ± 3 digits
660 nm, nominal
890 nm, nominal
< 15 mW
Fast
Alternating 1.5-second chimes
Flashing SpO
numeric or message
2
User-selectable high and low SpO
low pulse rate
Sensor errors, connection errors, insufficient signal,
communication problem, internal calibration error, or self-test
failure.
TM
250cx SpO
© 2013 by General Electric Company. All rights reserved.
Specifications
; User-selectable high and
2
parameter has not been demonstrated.
2
2