Hanna Instruments HI 98188 Instruction Manual page 36

Usp compliant ec, tds, nacl, resistivity, temperature meter
Table of Contents

Advertisement

The following table lists the specific range parameters:
Item
Description
Calibr.
Number of days after
Timeout
Calibration warning
(EC, NaCl)
is displayed
Out cal
Display warning if the
range check
reading is too far from the
(EC range
calibration points
only)
Temperature
Temperature
source
entry mode
Temperature
compensation
mode
Range select
Fix one
specific range
Cell constant
Manual set of
the cell constant
Temperature
Set the coeficient
coeficient
for linear temperature
compensation
Temperature
Reference temperature 15 °C, 20 °C, 25 °C
Ref
Temperature unit
TDS factor
Valid value
Default
Disable, 1 to 7 days
Disable
Enable/Disable
Disable
Probe/Manual
No TC, Linear
Non Linear
Automatic, Fix one
Automatic
resolution of
EC or resistivity range
0.010 to 10.000
1.000
0.00 to 10.00 %/°C
1.00 %/°C
°C or °F
0.40 to 1.00
36
Probe
No TC
25 °C
°C
0.50