Graph The Measurements - Fluke 2638A User Manual

Hydra series iii data acquisition unit
Hide thumbs Also See for 2638A:
Table of Contents

Advertisement

2638A
Users Manual
Statistic
Maximum
Minimum
Average
Standard Deviation
Peak-to-Peak
Rate of Change
To view the scan data:
1. Start a scan. See "Start a Scan" on page 4-10.
2. Push  to open the Scan Data menu. After the menu opens, push  to cycle the menu
pages to view more channels.
3. To view channel statistics, push .

Graph the Measurements

The Scan function has a graph feature that lets the user display the measurement data in a
graph. This feature is located in the Scan menu (push
automatically scales to fit all the information on the display when opened. To refine the
data, the graph can be manipulated with the front-panel arrow keys as shown in
Figure 4-5.
Selected
Channels
and Key
Select or Deselect
4-12
info@Fluke-Direct.com
Table 4-4. Scan Statistics
Maximum measurement.
Minimum measurement.
Average of all the measurements taken.
A measure of the distribution of a set of data from its mean. The more spread
apart the data, the higher the deviation.
Difference between the Maximum and the Minimum.
A measure of how much the readings change over time.
The maximum number of channels that can be graphed at once is four.
Selection of two or more channels with different input types (such as
temperature and milliamps) can make the graph difficult to read because
they do not share the same unit. To prevent this, only select channels of the
same input type or view the channels one at a time.
Figure 4-5. Graph Feature
Fluke-Direct
Description
then push ).The graph
Notes
Zoom Out
Zoom In
Move Left (in History Mode)
Move Right (in History Mode)
Time Stamps
Additional Setup Options
Graph Controls
(Front-Panel Keys)
hcn056.eps
.com
1.888.475.5235

Advertisement

Table of Contents
loading

Table of Contents