Review Memory

The Memory mode lets you view the stored images. You can also delete
images in this menu.
1.
Highlight the Memory icon.
2.
Push  to open the Memory mode.
3.
Push / to scroll through and review the stored images.
4.
Push  to delete image.

Emissivity

The emissivity is adjustable in 0.01 steps from 0.10 to 01.00. The default value
is set at 0.95.
The correct emissivity values are important for you to make the most accurate
temperature measurements. Emissivity of a surface can have a large effect on
the apparent temperatures that the Product observes. Understanding the
emissivity of the inspection surface can, but not always, allow you to obtain
more accurate temperature measurements.