Most laboratory detergents are used at recommended concentrations. Neutral detergents, such as
Liquinox, are safer to use when regular cleaning is necessary. To decrease the cleaning times,
increase the temperature or use an ultrasonic bath. To complete the cleaning, rinse a few times with
deionized water and then let the sample cell air dry.
Sample cells may also be cleaned with acid, followed by a thorough rinse with deionized water.
Note: Always use acid to clean sample cells that were used for low-level metal tests.
Special cleaning methods are necessary for individual procedures. When a brush is used to clean
sample cells, take extra care to avoid scratches on the interior surfaces of the sample cells.

Replace the batteries

Explosion hazard. Expired batteries can cause hydrogen gas buildup inside the instrument. Replace
the batteries before they expire and do not store the instrument for long periods with the batteries
installed.
Refer to
Install the batteries
Troubleshooting
Problem
Absorbance > 3.5!
Over range!
Under range!
Reading Failed!
Initializing Failed!
Calibration Needed!
Air Scan Needed!
Please attach the
module!
Failed in sending data!
Replacement parts and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
W A R N I N G
on page 7 for battery replacement.
Possible cause
The measured absorbance is more than 3.5.
The concentration is more than the upper limit
of the current method.
The concentration is less than the lower limit of
the current method.
There is an electronic or optical defect.
While initializing, the instrument failed.
The calibration data is corrupt.
The calibration data is corrupt.
The module is not found when data is sent.
W A R N I N G
Solution
Dilute the sample and repeat the
reading.
Dilute the sample and repeat the
reading.
Examine the sample.
Contact technical support.
Set the instrument to off and then
to on.
Contact technical support.
Contact technical support.
Contact technical support.
Insert a module.
Make sure that the module is
inserted and latched correctly.
English 19