MAN0530 Issue 10 - 11/09
2104M0506
APPENDIX A - HANDHELD INTERROGATOR
•
view the system type
•
set the system path length
•
perform a system self test
•
check system alignment
•
initialise the system and make it live
Note: The 4-20 output is inhibited during this procedure. See the earlier description.
system type
The initial display after Install is chosen is in the following format:
where:
tttttttt
Note: It is not possible to change the system type. This is set during calibration and configuration
of the unit at the factory.
Path length
The following message is displayed:
where:
nnnn
The following ranges of path length can be entered for the system types:
System Type
Short Range
Medium Range
Long Range
Duct Mount
System type. The system type displayed is one of the following:
Display Text
Short Range
Medium Range
Long Range
Duct Mount
The path length in metres.
Minimum Range (m)
5
20
120
0.5
System Type
tttttttt
System Type
40m range system
120m range system
200m range system
0.5 - 5m range system
Path Length
nnnn
Maximum Range (m)
40
120
200
5
88
Searchline Excel
Step Size (m)
1
1
1
0.1