UV LAMP PARAMETERS

Lighting UV lamp for 24 hours

If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, the fan cover, the duct, the fan, the
humidifying pan, and the humidifying pan cover. Clean all the attachments in an autoclave or with alcohol
for disinfection.
2. Clean and wipe off inside the chamber with alcohol for disinfection.
3. Set the CO
density to 0 %. Refer to page 32~33.
2
4. Press Menu key to lead the Menu screen.
5. Press Tools #1 key to lead the Tools #1 screen.
6. Press UV Setting key to lead the UV Setting
screen.
60