More transient examples
The previous examples showed how the transient system can be used to control the output voltage
amplitude. The transient system can also control output frequency, phase, waveform shape, voltage and
frequency slew rates, offset voltage, and peak current limit. The following examples illustrate how the
transient system's Pulse mode can generate frequency, shape, phase, and voltage slew pulses.
Freq
FREQ:M PULSE
FREQ 60
FREQ:T 50
Pulse
WIDTH .1
Trigger Control
INIT IMMED
Shift
Trigger
Shape
SHAPE:M PULSE
SHAPE SINE
SHAPE:T SQUARE
Pulse
WIDTH .05
Trigger Control
INIT IMMED
Shift
Trigger
Shift
Phase
PHASE:M PULSE
PHASE 0
PHASE:T 180
Pulse
WIDTH .05
Trigger Control
INIT IMMED
Shift
Trigger
Voltage
VOLT:M PULSE
VOLT 120
VOLT:T 150
SLEW:M PULSE
SLEW 10000
SLEW:T 1000
Pulse
WIDTH .1
Trigger Control
INIT IMMED
Shift
Trigger
Quick Start Guide
FREQ TRIG
FREQ
Trigger
SHAPE TRIG
SHAPE
Trigger
PHASE TRIG
PHASE
Trigger
SLEW TRIG
Trigger
width
width
width
SLEW
width
15