Type k j thermometer

Advertisement

    Command R:
Equivalent to one pushing on the REL button and no message is
returned.
    Command C:
Equivalent to one pushing on the / button and no message
is returned.
    Command A:
nd
1
BYTE:
The first byte is the start byte , it value is 2.
nd
2
BYTE:
bit7
bit6
bit5
bit4
bit3
C/F
Low Bat
Hold
REL
bit 2
bit 1
bit0
0
0
0
→ normal mode
0
0
1
→ MAXIMUN mode
0
1
0
→ MINIMUN mode
1
0
0
→ AVG mode
1
1
1
→ calculate MAX/MIN/AVG in back-ground
and lcd "MAX""AVG""MIN" will flash.
bit3: 1→0→K TYPE 1→J TYPE
bit4: 1→REL
bit5: 1- HOLD 0→not HOLD
bit6: 1→LOW BATTERY 0→BATTERY NORMAL
bit7: 1→ 0→
3
nd
BYTE:
bit7
bit6
bit5
bit4
bit3
point
minus
OL
bit0:1→main window value is OL, 0→not OL
bit1:1→main window value is minus, 0→main window value is plus.
bit2:1→4
th
byte and 5
th
byte represent #### 0→4
Type K J Thermometer
bit2
bit1
bit0
K/J
MAX/AVG/MIN
bit2
bit1
bit0
point
minus
OL
th
byte and 5
th
byte

Advertisement