Mechanical Specifications; Electronic Specifications - Omega Engineering FPR 200 User Manual

Flow sensor with polypropylene body flow sensor with stainless steel body
Table of Contents

Advertisement

Mechanical Specifications

FLOW SENSOR WITH STAINLESS STEEL BODY
Maximum pressure:
Maximum temperature: 225 º F
Measuring range*
1/2" NPT Ported:
3/4" & 1" NPT Ported: 1.5 - 50 GPM
Wetted materials:
Non-wetted materials: Epoxy, Lexan
FLOW SENSOR WITH POLYPROPYLENE BODY
Maximum pressure:
Maximum temperature: 150 º F
Measuring range*
1/2" NPT Ported::
Wetted materials:
Non-wetted materials: Epoxy, Lexan
*Important: Chose a maximum flow rate. For 1/2": 5 - 15 maximum GPM.
For 3/4" and 1": 15 - 50 maximum GPM. Minimum flow rate will
be 10% of maximum flow rate. Example: If your maximum flow
rate is 8 GPM, the minimum flow rate would be .8 (8 x .1 = .8).
Thus, the correct flow range would be .8 - 8.0 GPM.

Electronic Specifications

ALL VERSIONS WITH 4-20 MA CIRCUIT
Power requirements:
Load driving capacity:
Trans. Distance:
LED:
Flow rate accuracy:
Repeatability:
Resolution:
Response time:
Over-current protection: self limiting at 30mA
Fail safe indication:
200 PSI clear cover, 500 PSI SS cover
0.5 - 15 GPM
316SS, acetal copolymer, Buna, PEEK
and polycarbonate
®
and PVC
150 PSI
0.5 - 15 GPM
316SS, acetal copolymer, Buna, PEEK,
polycarbonate and polypropylene
®
and PVC
12-35V
, 20mA
DC
Max. Load (
) = 50(Power Supply Voltage - 12)
limited only by wire resistance & supply voltage
provides power indication
±2% of full scale
±0.5% of scale
infinite
2 seconds to 90%
4 mA
(11)