Adjusting Fork Assembly Height
NOTE: For this procedure, refer to FIGURE 5. 8 and 5.11.
1. Remove the dust cover from the Caster head tube.
2. Unscrew the Lock nut (M12) that secure the Vertical axis assembly to the Caster head
tube.
3. Replace the Small spacer (10 mm) by a Large spacer (47, 5 mm) and a new longer
caster vertical axis if required. Refer to table Seat-To-Floor-Height, page 35.
4. Secure the caster vertical axis assembly to the Caster head tube.
5. Do not overtight the locknut (10-13 Nm), you may RISK to damage the bearings and
the axis.
6. Repeat STEPS 1 to 5 for the other caster assembly.
NOTE: Both fork assemblies must be set at the same height.
Small spacer
Position 1 = 10 mm
Large spacer
Position 2 = 47,5 mm
F I G U R E 5 . 1 1 Adjusting Fork Assembly Position
SECTION 5—REAR WHEELS/FRONT CASTERS
Dust cover
Caster
head tube
Caster vertical
axis short or
long (M12)
3 3
Lock nut
(M12)
Frame