Nokia RM-235 Service Manual page 84

Table of Contents

Advertisement

4.
Choose Tuning -> Ambient Light Sensor Calibration. You should see the following window:
5.
Read AD-count values for Channel 0 and Channel 1 by click Read button and write them down.
6.
Repeat 1-5 for the phone to be calibrated and make sure the phone to be calibrated is located in the
same place as reference phone was when luminance reading was taken.
7.
Calculate co-efficient from reference phone and phone to be calibrated AD-count values by division: Co-
efficient = AD-count(reference phone) / AD-count(phone to be calibrated), write down the calculated co-
efficient values.
8.
-> Iterate by changing Channel 0 and Channel 1 (reference level) values (remove cross from 'Use default
values only'). After writing some value to Channel 0 and Channel 1 (reference value), calibrate button
must be pressed. Stop iterating when Co-efficient is equal to Co-efficient calculated in step 7. Note that
decimal numbers should be used in the iteration in order to achieve enough precision (e.g. 200.2455)
9.
After having same Co-efficient value in "Co-efficient" textbox as the calculated value, make sure that
ambient light values (read using Testing → Display Test → "Luminance" textbox) are almost the same
in reference phone and calibrated phone. Remember that illuminance readings for reference and
calibrated phones must be done in the same ambient light conditions. If illuminance values differs a lot
(difference max. +- 10%), repeat whole ALS re-tuning procedure.
10. To end the calibration, click Close.
Page 3 –36
COMPANY CONFIDENTIAL
Copyright © 2008 Nokia. All rights reserved.
RM-235
BB troubleshooting
Issue 1

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

N78L4L3

Table of Contents