1-9-3. Recovery Method for Kernel/FPGA Version Upgrade Failure (this unit)

When the kernel/FPGA version upgrade using Memory Stick is failed, and the power does not turn on,
there is a possibility of recovery by writing the kernel/FPGA data using the writing tool.
Required equipment
. Personal computer (PC)
(equipped with USB port and LAN port)
. RS-232C, D-Sub 9 pin straight cable (female-female)
. USB 5 RS-232C conversion adaptor
n
When PC has RS-232C terminal, this is not required.
. Application tool
Name
BVM\HAB_TOOL ver1.1
Preparation
1. Remove the rear cabinet. (Refer to Section 1-3-1.)
2. Set the switch (S7012) on the B board to Center.
3. Set "1" of the DIP switch (S7000) on the B board to
ON.
4. Connect PC with this unit using the RS-232C cable.
1-48
Copyright
Reference information
Sony monitor
If it does not start, "NET Framework" may be required.
Download "NET Framework version 2.0" from the Microsoft
download center. Refer to "Readme.text" in the folder.
OFF
4 3 2 1
ON
L
C
S7000/B board
S7012/B board
R
BVM-L230