Welcome to the CareSens N Voice Blood Glucose Monitoring System Thank you for choosing the CareSens N Voice Blood Glucose Monitoring System. The system provides you with safe, fast, and convenient blood glucose in vitro (i.e., outside the body) monitoring. The bilingual talking function of CareSens N Voice Meter can be used as an aid for using the meter.
Table of Contents Information Important Information: Read This First! Specifications CareSens N Voice Blood Glucose Monitoring System Inserting or Replacing Batteries CareSens N Blood Glucose Test Strip CareSens N Voice Blood Glucose Meter CareSens N Voice Blood Glucose Meter Display...
CareSens N Voice • T he CareSens N Voice Blood Glucose Meter is designed to Blood Glucose Monitoring System should be used only for minimise code related errors in monitoring by using the self-testing outside the body (in vitro diagnostic use only).
5 seconds * Carrying Case * Batteries Sample type Fresh capillary whole blood Calibration Plasma-equivalent CareSens N Voice BGM System may include the following items: * CareSens N Blood Glucose Test Strips Assay method Electrochemical * Lancets Battery life 1,000 tests...
Inserting or Replacing Batteries CareSens N Blood Glucose Test Strip The CareSens N Voice Meter uses two AAA alkaline batteries. The CareSens N Voice Blood Glucose Monitoring System Before using the meter, check the battery compartment and measures blood glucose quickly and accurately. It automatically insert batteries if empty.
The cable for data management software can be young children. The test strip vial contains drying agents that ordered separately. Please contact Spirit Healthcare 24/7 are harmful if inhaled or swallowed and may cause skin or eye Freephone Support Line at 0800 881 5423 or go to www.spirit- irritation. healthcare.co.uk. 10 www.i-sens.com...
CareSens N Voice Blood Glucose Meter Display Setting Up Your System You should check and update the meter-settings such as time appears when the time alarm : and date before using your meter or after changing the meter alarm has been set batteries.
Page 9
“Set time 12-hour(24 hour)”. Press ▲ or ▼ button to select a format. The "AM/PM" is not displayed in the 24-hour format. After selecting the format, press the S button to progress to the next step. 14 www.i-sens.com www.spirit-healthcare.co.uk 15...
3 seconds to confirm all your selections and exit from SET mode. If the talking function is on, the meter will announce the set date and time for your confirmation. 16 www.i-sens.com www.spirit-healthcare.co.uk 17...
You may do a control solution test: • W hen you want to practice the test procedure using the control solution instead of blood • W hen using the meter for the first time 18 www.i-sens.com www.spirit-healthcare.co.uk 19...
Page 12
When the "ctl" symbol is Note: The CareSens Control Solution can be obtained free displayed, the result is stored in the meter's from Spirit Healthcare on 0800 881 5423. memory but is not included in the averages. 20 www.i-sens.com www.spirit-healthcare.co.uk 21...
You may use the lancing device contained in the The test result of each control solution should be within the CareSens N Voice Blood Glucose Monitoring System or any range printed on the label of the test strip vial. Repeat the other medically approved lancing device.
Note: The skin depth to get blood samples will vary for various people at different sample sites. The lancing device's adjustable tip allows the best depth of skin penetration to get an adequate sample size. 24 www.i-sens.com www.spirit-healthcare.co.uk 25...
3 seconds again before applying the blood sample or when the test result is displayed just after testing. Marking Post-meal Test Results The CareSens N Voice meter allows you to mark a result of a Applying Blood Sample post-meal test with symbol. The post-meal test mark ( ) Step 9 can be attached just before applying the blood sample.
Page 16
The meter may switch off, if the control solution is not applied within 2 minutes of the symbol appearing on the screen. If the meter turns off, remove the strip, reinsert, and start from step 2. 28 www.i-sens.com www.spirit-healthcare.co.uk 29...
The lancet is for single use only. Never share or reuse pressure. Keep holding the lancing device against your skin a lancet. Always dispose of lancets properly. until sufficient (at least 0.5 μL, actual size: ) blood is drawn. Carefully lift the lancing device away from your skin. 30 www.i-sens.com www.spirit-healthcare.co.uk 31...
Page 18
If the sample drop of blood runs or spreads due to contact with hair or with lines on your palm, do not use that • W hen insulin has the biggest effect sample. Try puncturing again in a smoother area. • T wo (2) hours after an insulin injection 32 www.i-sens.com www.spirit-healthcare.co.uk 33...
Note: Please contact the Spirit Healthcare 24/7 Freephone 'Pc' is displayed when the data cable connects Support Line at 0800 881 5423 or go to www.spirit-healthcare. the meter with computer. For more information, co.uk if such messages are displayed even if you do not have contact the Spirit Healthcare 24/7 Freephone hyperglycemia or hypoglycemia.
Meter Memory The CareSens N Voice Meter can save up to 500 test results with Step 3 Viewing Pre-meal Averages temperature, time and date. If the memory is full, the oldest test Repeatedly press the ▲ ▲button result will be deleted and the latest test result will be stored.
Setting the Alarm Function Four alarms can be set in the CareSens N Voice Meter: one post- Viewing Test Results meal alarm (PP2 alarm) and three time set alarms (alarm 1-3). The Step 1 PP2 alarm goes off 2 hours after setting the alarm. The alarms Press the S or ▲...
Press and hold the S button for 3 seconds to finish and turn the meter off. Press ▲ to adjust the hour of "alarm 1". A number representing the hour will blink on the screen. Press ▼ to set the hour. Press ▲ to end. 40 www.i-sens.com www.spirit-healthcare.co.uk 41...
“Error 4. The The blood Repeat the test sample was sample has using a new not enough abnormally test strip. or too thick.” high viscosity or insufficient volume. 42 www.i-sens.com www.spirit-healthcare.co.uk 43...
What to do "Error 6 / Error 7 There is a problem Do not use the / Error 8 / Error 9. with the meter. meter. Contact the Meter failure." Spirit Healthcare customer service. Note: If the error messages persist, contact the Spirit Healthcare 24/7 Freephone Support Line at 0800 881 5423 or go to www.spirit-healthcare.co.uk. 44 www.i-sens.com www.spirit-healthcare.co.uk 45...
System has been evaluated in laboratory and in clinical tests. • C heck whether the test strip is Accuracy: The accuracy of the CareSens N Voice System (Model inserted with the contact bars facing GM505UAA) was tested by comparing blood glucose results up.
Precision: Precision studies were performed in a laboratory Manufacturer's Warranty using the CareSens N Voice BGMS. i-SENS, Inc. warrants that the CareSens N Voice Meter shall be free of defects in material and workmanship in normal use for Within Run Precision a period of 5 years.