Display
Action
Not.enough.blood.or.control.solution.was.drawn.into.the.test.strip.for.measurement.or.
was.applied.after.the.test.had.started..Discard.the.test.strip.and.repeat.the.blood.
glucose.or.control.test.
Blood.or.control.solution.was.applied.to.the.test.strip.before.the.flashing.drop.symbol.
appeared.on.the.display..Discard.the.test.strip.and.repeat.the.blood.glucose.or.control.
test.
An.electronic.error.occurred,.or.in.rare.cases,.a.used.test.strip.was.removed.and.
reinserted..Turn.the.meter.off.and.on,.or.take.the.battery.out.for.20 seconds.and.
reinsert.it..Perform.a.blood.glucose.or.control.test.
The.temperature.is.above.or.below.the.proper.range.for.the.system..Refer.to.the.test.
strip.package.insert.for.system.operating.conditions..Move.to.an.area.with.the.
appropriate.conditions,.wait.5 minutes,.and.repeat.the.blood.glucose.or.control.test..
Do.not.artificially.heat.or.cool.the.meter.
The.battery.is.almost.out.of.power..Change.the.battery.now..If.the.message.reappears.
after.the.battery.has.been.replaced,.remove.the.battery.again,.press.any.meter.
button,.then.reinsert.the.battery..
The.time.and.date.settings.may.be.incorrect..Make.sure.the.time.and.date.are.correct.
and.adjust,.if.necessary..