GE AMX-4+ Technical Manual page 109

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
8 2
Handle Sensor Installation
Note:
7. Slide end cap along handle away from end casting.
8. Remove end casting from handle assembly by removing two hex socket head
capscrews securing it to handle springs. Be careful not to damage sensor and
drive enable switch leads.
9. Remove two binding head capscrews securing sensor to standoff posts inside the
end casting.
10. Pull sensor leads through back of end casting and remove sensor.
This procedure covers installation of either the right or left sensor in the han
dle assembly.
1. Position sensor on standoff posts inside end casting and secure with two binding
head capscrews. See Illustration 8-3.
2. Feed sensor leads through back end of casting to the connector.
3. Position handle assembly in end casting and secure handle springs to casting with
two hex socket head capscrews. Be careful not to damage sensor and drive enable
switch leads.
4. Slide end cap along handle and position on end casting.
5. Secure end cap on end casting with four hex socket button head capscrews.
6. Connect sensor leads (red, white and black - positions 3, 4 & 5 in connector) to
connector). See Illustration 8-2.
7. Fasten handle assembly to two support brackets with eight hex flange head
capscrews. See Illustration 8-1.
8. Connect connectors at each end of handle assembly.
9. Connect ground wire, located at left end of handle, to mainframe.
10. Attach wires to the handle assembly angle support bracket with ty-raps.
11. Close top cover. Refer to Section 5-6.
8-3
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2169360 series2236420 series2275938 series

Table of Contents