GE AquaTrans AT868 User Manual page 51

Ultrasonic flow transmitter for water (1- & 2-channel)
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

September 2004
Selecting Error Handling -
ERROR (cont.)
Table 2-5: Error Options and Responses for a 2-Channel/Path Meter
When Measuring
Display Response
CH1 or CH2
Holds last "good"
(vel, vol, etc.)
reading.
Adds two channels/
SUM
paths using the last
"good" reading.
Subtracts two channels/
DIF
paths using the last
"good" reading.
AVE
See Error Handling for Average Measurements below.
2-26
Table 2-4: Error Options and Responses for a 1-Channel Meter
Option
Output Response
HOLD
Holds last "good" reading.
NO UP
Totalizer Response When Error Handling is
Holds last "good" reading
and continues to totalize
based on that "good" reading.
Holds last "good" reading
and continues to totalize
based on two channels/paths.
Holds last "good" reading
and continues to totalize
based on two channels/paths.
7. The ERROR-HANDLING D display appears. This allows the user to
set a time delay before the instrument reports that an error has
occurred. Enter the desired time delay in seconds by using the
arrow keys. Press [
[ ] and [ ]
location and press
number and press [ENTER].
8. Do one of the following:
For a 1-channel meter, go to Procedure Options on page 2-27.
For a 2-channel/path meter, proceed to Step 9 on page 2-27.
Totalizer Response
Holds last "good" reading
and continues to totalize,
based on that reading.
Stops totalizing.
HOLD
LOW, HIGH, HHIGH
Stops totalizing.
Stops totalizing if
either or both channels/
paths go into error.
Stops totalizing if
either or both channels/
paths go into error.
]
and [ ] to move the cursor to desired
to increase or decrease the
Programming Site Data