Install The Exhaust Tank - Panasonic MLS-3751 Operating Instructions Manual

High-pressure steam sterilizer
Table of Contents

Advertisement

How to Use

1. Install the exhaust tank.

1) Take out the exhaust tank from the product and remove the
packings.
Advisory note
Do not pull out the exhaust tank by force from the product with the
exhaust hose being connected. Otherwise, trouble may be caused.
2) Pour water up to the LOW (minimum) level mark, insert the packings
to the holes of the product and connect the exhaust hoses.
Memo
The water in the exhaust tank increases in each sterilizing operation.
If it will increase up to the HIGH (maximum) level mark, once discard
the water and fill the tank with fresh water up to the LOW level mark.
3) Make sure that the exhaust hose packings have been
inserted completely in the holes of the exhaust tank.
When using the exhaust tank attached to the product,insert
the exhaust hose packings securely in the holes of the
exhaust tank.
Be sure to use the exhaust tank attached to the product.
If the packings of the exhaust hose have not been inserted
sufficiently, steam may escape and result in burns.
Be careful about the position of the exhaust hose
insertion holes.
Be sure to insert the hose in the tank into the holes at
the back inside the exhaust tank.
Failure to observe this caution may cause large steam leaks
from the tank on the rear side.
Advisory note
Insert the packings up to the hilt while turning them clockwise and
counterclockwise. If they are turned in only one direction, the exhaust
hoses will be twisted to cause trouble.
4) Stow the exhaust tank inside the product.
Make sure that the exhaust hoses are not bent or twisted. If they
are so, air in the chamber is hard to escape and an error message
may appear.
16
CAUTION
( Continued )
HIGH level mark
LOW level mark
Exhaust hoses
Packing
Hole
Hose in tank
Exhaust tank

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mls-3751lMls-3781Mls-3791l

Table of Contents