Connect And Adjust The Test Instruments - GE Bently Nevada 3300/25 Operation Manual

Dual accelerometer input monitor
Table of Contents

Advertisement

15.2 Connect and Adjust the Test Instruments

(Repeat for channels A and B.)
1.
Connect the function generator to the channel A terminals as shown in the figure
above.
(Note: For a single transducer input system, do not repeat this step for channel B.)
2.
Adjust the function generator sine wave to 307.5 ± 2 Hz with a -7.5 ± .3 Vdc offset.
3.
Adjust the signal amplitude to the meter full scale according to Full Scale option
(BB for Channel A, CC for Channel B) as shown in the table:
SIGNAL
OPTION
AMPLITUDE
(mV peak)
01
200
02
500
03
1000
04
2000
05
500
06
1000
NOTE: If barriers are used, calibrate with barriers in place.
SIGNAL
AMPLITUDE
OPTION
(mV RMS)
141
11
354
12
707
13
1414
14
354
15
707
16
17
Section 15 - Calibrate Channels
SIGNAL
AMPLITUDE
AMPLITUDE
(mV peak)
(mV RMS)
204
510
1020
2039
492
985
1970
SIGNAL
144
361
721
1442
348
697
1393
39