8-5-19-5
Main Cable Cover installation
1.) Install the Main Cable Cover so the Column Cover Assembly overlaps Main Cable Cover edges
(see:
2.) With the console raised to its full height, install the 3 screws to secure the Main Cable Cover and
Column Cover Assembly.
3.) Lower the console and install the lower right side screw (see:
lower right side screw placement" on page
4.) Position the lower Column Cover tab inside the Front Cover.
5.) Install the Top Cover.
6.) Install the Side Covers.
8 - 78
. 2
EV
Figure 8-33 "Main Cable Cover (Column Cover overlaps Main Cable Cover)" on page
Figure 8-34 Position the column cover tab inside the front cover
Section 8-5 - Replacing Covers and Bumpers
Figure 8-29 "With console lowered,
8-73).
LOGIQ E9 S
M
ERVICE
ANUAL
8-77).