Publication Information The information in this manual applies only to T2100-ST1 Treadmill, and T2100-ST2 Treadmill,. It does not apply to earlier versions. Due to continuing product innovation, specifications in this manual are subject to change without notice. T2100-ST1, T2100-ST2, CASE, CardioSoft, and MAC are trademarks owned by GE Medical Systems Information Technologies, Inc., a General Electric Company going to market as GE Healthcare.
Incline Range ..........................24 Running Surface ........................24 Communication Ports ......................25 Floor Surface Footprint ......................25 Operating and Storage Condition Recommendations ..........25 Power Requirements ......................25 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Page 4
Elevation Screw Lubrication ....................43 Running Deck Maintenance ....................43 Troubleshooting ....................45 Power Switch Will Not Illuminate ..................45 Facility Circuit Breaker Trips when Powering Up ............46 T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Page 5
Treadmill Will Not Start ......................46 Running Belt Slips When In Use ..................46 Running Belt Is Off-Center ....................46 Internal Circuit Breaker Location and Resetting ............46 Maintenance Log ....................47 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Indications for Use The T2100-ST1 and T2100-ST2 treadmills are designed for cardiac stress testing. Prescription Device Statement CAUTION: United States federal law restricts this device to sale by, or on the order of, a physician.
The following messages apply to the system as a whole. Specific messages may also appear elsewhere in the manual. WARNING: The T2100-ST1 and T2100-ST2 treadmills are manufactured to exacting standards both in physical form and in component selection. The components used in our products have been selected with performance and medical safety in mind.
Page 9
Ensure the master power switch is in the off position before plugging in the T2100-ST1 or T2100-ST2. A power surge could damage the sophisticated electronic system of the treadmills. WARNING:...
Degree of protection against Type B external application applied electrical shocks part. Degree of protection against harmful Ordinary equipment (enclosed ingress or water equipment without protection against ingress of water). T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Mode of operation Continuous operation. Regulatory and Safety Conformance T2100-ST1 and T2100-ST2 treadmills meet the following safety and regulatory standards for FDA Class 1 motor operated physical medicine machines. They have been tested by Intertek Testing Services N.A Inc., and are listed by Engineering Testing Laboratories (ETL).
Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized GE Healthcare. The electrical installation of the relevant room complies with the requirements of the appropriate local, state, and other government regulations.
Page 13
Check input communications cord connection (if equipped) for proper interface with all equipment. Keep all cords clear of patient to avoid trip hazards. 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Any hazard symbols on your device or packaging that is in black and white indicates a potential hazard and is a caution. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Page 15
Customers should contact their local authorities or T2100 Distributor for guidance in complying with the directive. 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Page 16
Device is suitable for the external application of the type “B” applied parts. Label Locations This section identifies the labels and their locations on the product and packaging. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Page 17
Package information for shipping: Model number Reference number Serial number Storage conditions Regulatory compliance Country of origin EC Representative information 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Service Information This section provides information pertaining to the maintenance and servicing of the system. Familiarize yourself with this information before requesting service from GE Healthcare or its authorized representatives. Service Requirements Failure on the part of the responsible individual, hospital, or...
GE Healthcare service personnel should service the device. Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user's responsibility to report the need for service to GE Healthcare or to one of their authorized agents. Additional Assistance...
Related Documents The following documents are referenced in this manual and provide additional information that may be helpful in the installation, configuration, maintenance, and use of this product. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
If you have not received training on the use of the system, you should request training assistance from GE Healthcare. To see available training, go to the GE Healthcare training Web site (http://www.gehealthcare.com/usen/education/index.html). Under the Technical Service Education section, select Diagnostic Cardiology.
Product Overview The T2100-ST1 and T2100-ST2 treadmills are designed and built to withstand the extraordinary demands of medical devices and are compatible with CASE, CardioSoft/CS, MAC 5500, and MAC 2000. References to left, right, front, and rear are based on the assumption that you are standing on the treadmill, facing the handrails.
All steel construction with powder-coat finish Treadmill net weight: 425 lb., 193 kg Drive System Heavy-duty 6-peak hp. brushless, DC servo motor T2100-ST1 110-120VAC, 1-phase, 60 Hz, 20-amp power supply T2100-ST2 220/240VAC, 1-phase, 50-60 Hz, 15-amp power supply Speed Range 0.1 to 15.0 mph, 0.2 to 24.0km/h, self-calibrating and adjustable in...
5% per each additional 500 feet (152m) above 5,280 feet (152m) Power Requirements The T2100-ST1 is designed to operate on a dedicated 110-120 VAC 20-amp power supply and the T2100-ST2 is designed to operate on a dedicated 220/240 VAC 15-amp. Make sure that the treadmill is connected to an outlet that looks like the following illustration.
Ensure the master power switch is in the off position before plugging in the T2100-ST1 or T2100-ST2. A power surge could damage the sophisticated electronic system of the treadmills. NOTE: The T2100-ST1 and T2100-ST2 Treadmills must have their own dedicated power outlet.
If your treadmill must pass through a door opening less than 36” wide, additional disassembly will be required. This task should be performed by an authorized service provider to ensure that the T2100-ST1 and T2100-ST2 are properly reassembled and functioning correctly. Safe Handling Guidelines ...
Assembly and Setup WARNING The T2100-ST1 and T2100-ST2 treadmills each weigh 425 lbs. When lifting the rear of the treadmill and rolling it on its front wheels, you are moving 132 lbs. This can be accomplished by one person, but each installer must evaluate whether he or she is capable of moving this amount of weight without causing strain or injury.
When folding handrails apply cardboard between frame and handrail to prevent handrail damage. Location Place the T2100-ST1 or T2100-ST2 on a firm and level hard surface that is free of tile grout lines. The illustration below shows the minimum recommended clearances from the treadmill edges to any obstruction for dismount and safety purposes.
Belt Tracking System significantly ® reduces the need to adjust the belt on your T2100-ST1 or T2100-ST2 treadmill. However, when you operate your treadmill for the first time, you may need to adjust the tracking of the belt to conform to your floor.
Final Setup - Test Plug Procedure Each T2100-ST1 and T2100-ST2 treadmill includes an RS-232 test plug that enables you to test the operation of the treadmill without the ECG unit attached. The plug is located on the left side of the treadmill secured to the frame by Velcro .
NOTE: Unsuccessful completion of the preceding testing procedure indicates a problem with the setup. Call GE Medical Systems Information Technologies to trouble shoot failure of test plug procedure. Remove the test connector.
RS232 port and a USB Type B port offer connectivity diversification. To communicate to the T2100-ST1 or T2100-ST2 treadmill using the USB port, you will need to install the appropriate USB driver software on your host computer. The USB driver is supplied on the provided flash drive.
Operating Instructions Before operating the T2100-ST1 or T2100-ST2, familiarize yourself with the following control safety features of the treadmills. Electrical Safety Tests The electrical safety of this installation is the responsibility of the customer, not GE Medical Systems Information Technologies. In hospitals, contact your in-house biomedical technician, electrician, or technically qualified personnel.
To re-attach the pull tether, attach clip to the original position on the switch. The treadmill will return to 0.0% elevation. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Adjust the handrail to the desired height. Release the adjustment knob to secure the handrail in the current position. Turn the adjustment knob clockwise twice to lock the handrail. 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Page 38
Operating Instructions T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Contact your local GE customer support representative. Use only authorized replacement parts. Using other parts may void your warranty and may cause your T2100-ST1 or T2100-ST2 treadmill to malfunction. Daily Maintenance Wipe the treadmill to remove soil, moisture, and perspiration.
When belt is clean, stop the treadmill. Inspect the running belt for tears or nicks. If damaged, replace the belt. Perform the Running Belt Tracking Adjustment and Belt Tension Adjustment. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
The patented MasterTrack Belt Tracking System significantly ® reduces the need to adjust the belt on your T2100-ST1 or T2100-ST2 treadmill. However, when you operate your T2100-ST1 or T2100- ST2 treadmill for the first time, you may need to adjust the tracking of the belt to conform to your floor.
NOTE: Because this adjustment is not covered under your warranty, refer to "Drive Belt Tension Adjustment" in the Preventive Maintenance chapter of the T2100-ST1 / T2100-ST2 Field Service Manual. NOTE: If there are indications that the drive belt has stretched and...
Preventive Maintenance Exterior Care The powder-coat finish on your T2100-ST1 or T2100-ST2 treadmill is an extremely durable finish and requires minimal care. Do not allow perspiration to build up on your treadmill. Wipe the unit daily. Elevation Screw Lubrication The Elevation Screw must be cleaned and lubricated every 6 months to maintain proper operation of the treadmill.
Your in-house biomedical technician or technically qualified personnel can perform basic troubleshooting. Advanced troubleshooting may be found in the T2100-ST1/T2100-ST2 Service Manual shipped with your system, or contact GE Medical Systems Information Technologies, Inc., and open a customer-billable service call.
2 ea. Elevation Motor (3 amp [110v or 220v]) To reset a breaker, push the button. NOTE: Contact GE Medical Systems Information Technologies, Inc., for information on possible causes of tripped circuit breakers. T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V...
Maintenance Log Serial # ________________ Date Purchased____________ DATE HOURS SERVICE COMPLETED COST 2097937-001 Rev E T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 12 June 2017...
Maintenance Log T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V 2097937-001 Rev E 12 June 2017...
Page 50
GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223, USA Tel: +1 414 355 5000 1 800 558 5120 (US Only) GE Medical Systems Information Technologies, Inc., a General Electric Company, going to market as GE Healthcare. www.gehealthcare.com...