Errors That May Not Be Errors; Tips - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-18-7

Errors That May Not Be Errors

No DVR found in system – OK if there is not a DVR!
Info
\\192.0.2.178 /user:root ""] ExitCode=2: GetLastError returned 0Info
EchoVcr.GEdvrRecordUnit(3716); Waiting till DVR board become
alive...Error
No PatIO found in system – OK if there is not a Patient I/O!
Error
7-9-18-8

Tips

If the application does not allow the use of ALT-D (crashes quickly or will not start) the desktop icon
"GatherLogs" performs the same function.
-
Search for "Starting log system" to find start of session.
Logfile.txt only grows to ~100MB then copied to logfile.txt.bak.
Image Quality issues usually do not show up in log files.
Log files do not show everything.
PROPRIETARY TO GE
. 7
EV
; EchoVcr.GEdvrRecordUnit(3716); Failed to start process[net use
; EchoVcr.GEdvrRecordUnit(3716); DVR Board is not found!!
; PatioUsb(2300); OpenDriver Failed; 2
In extreme circumstances, reload the C: partition only. This allows access to the desktop and
use of "GatherLogs".