Page 2
TR-Series Thank you for purchasing the Fast Weigh MS-600 digital pocket scale. Please read all operating instructions carefully before use. Scales are precision measuring devices and should always be handled with proper care. To ensure years of reliable service, keep these simple tips in mind: •...
Page 3
I. Part List 1. Scale 2. Cover / Weighing Tray 3. AAA Batteries (x2) II. Key Description Weight Indicator ON/OFF/ Mode Key Tare Key III. Battery Installation 1. Before using the scale for the first time, check the battery compartment and remove any plastic insulation strips or packaging that may be in place to prevent battery drainage during shipping/storage.
Page 4
2. If the scale is turned off then back on, it will default to the last unit used. • TR-100 Units: g (gram) oz (ounce) ct (carat) gn (grain) • TR-600 Units: g (gram) oz (ounce) ozt (troy ounce) dwt (pennyweight) VI.
Page 5
2. If the displayed value is not within acceptable tolerance (see chart below), adjustment is required. Model Acceptable Calibration Recommended Tolerance (±) Weight Accuracy Class TR-100 0.03g 100g OIML M2 TR-600 0.2g 500g OIML M2 VIII. Adjustment 1.
Page 6
TR-Series IX. Troubleshooting • Problem: Scale will not turn on Solution: Check the batteries • • Problem: Displayed weight fluctuates randomly • Solution: Make sure your workspace is stable and free of any air currents or vibration Problem: Displayed weight is inaccurate • •...
Page 7
XI. Specifications TR-100 TR-600 Max. Capacity 100g 600g Readability 0.01g 0.1g Min. Weight 0.03g 0.2g Power 2 x AAA Alkaline Batteries Dimensions 0.8 x 2.4 x 3.8” Platform Dimensions 2.0 x 2.3” Warranty 5 Year Limited Warranty XII. Capacity/Readability Charts TR-100 100g x 0.01g...