Wixey WR510S DRO
Calibration
#CALWR510S

Advertisement

loading

Summary of Contents for Oliver Wixey WR510S DRO

  • Page 1 Wixey WR510S DRO Calibration #CALWR510S...
  • Page 3 STEP 1: Plane a board with a flat bottom to 1-1/4” or 1”. STEP 2: Use a caliper to measure the thickness and adjust the scale, if needed, as this will be your reference when calibrating the WR510S DRO Scale WR510S DRO Model 4430 Shown...
  • Page 4 Adjust WR510S display to the same height/number as the scale STEP 4: Loosen this screw to move the DRO to 6mm on the scale and tighten the DRO STEP 3: Lower the upper casting to its lowest position which is 6mm on the scale...
  • Page 5 Setting WR510S STEP 6: Press & hold【ON/OFF/CAL 】button When ABS blinks, release the【ON/OFF/CAL 】button, and it should become zero. STEP 5: Press【ON/OFF/CAL】button to start the display ABS mm mode STEP 8: Setting numbers to 6mm: press 【+】 each time to increase 0.1mm while press 【+】increases 0.1mm 【- 】each time to STEP 7: With ABS is still blinking decrease 0.1mm.
  • Page 6 After calibration, lift the upper casting to 6” (10014) or 8” (4430) and check if the scale and DRO show the same. SCALE...