Battery powered digital pressure gauges w. memory options & selectable units

Advertisement

Quick Links

Cecomp
Battery Powered Digital Pressure Gauges w. Memory Options & Selectable Units
®
Ranges and Resolution
See table below for standard ranges and units
Resolution is fixed for each engineering unit
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Standard accuracy: ±0.25% of full scale ±1 least significant digit
HA accuracy option: ±0.1% FS ±1 LSD, see ranges for availability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric
BL models: red LED backlight
Batteries, Battery Life
B:
2 AA alkaline, approx. 2000 hours
BL:
2 AA alkaline, approx. 150 to 1500 hours depending on
backlight usage
Controls and Functions
Three button keypad: Zero/clear, on/off, memory
BL models: Backlight active for 1 minute (user configurable)
F20B Peak Reading Memory
-M4: MEM 1 ~ MEM 4 or 4 tire designations: LF, RF, LR, RR
-M8: MEM 1 ~ MEM 8 or 4 tires plus 4 inner liner designations:
LF, LF IN, RF, RF IN, LR, LR IN, RR, RR IN
F22B Min/Max Memory
Minimum and/or maximum readings stored in memory, readings
cleared or stored at shutoff. User configurable.
Auto Shutoff
User selectable 1 minute to 8 hours or front button on/off
Factory default 5 minutes, unless other time is specified
Sensor Ranges and Engineering Units
3 psig ‡
Res
15 psig vac ‡
3PSIG
.001 100KPAVAC
6INHGG
.001 0.1MPAVAC
85INH2OG
.1
1BARVAC
50ZING
.01
1KGCMVAC
210GCMG
.1
1ATMVAC
150MMHGG
.1
15 psig
150TORRG
.1
15PSIG
200MBARG
.1
30INHGG
200CMH2OG
.1
400INH2OG
2000MMH2OG
240ZING
1
7FTH2O
.001 1000GCMG
20KPAG
760MMHGG
.01
5 psig ‡
Res
760TORRG
5PSIG
.001 1000MBARG
10INHGG
.01
1000CMH2OG
140INH2OG
.1
35FTH2O
80ZING
.1
100KPAG
350GCMG
.1
0.1MPAG
260MMHGG
.1
1BARG
260TORRG
1KGCMG
.1
350MBARG
1ATMG
.1
350CMH2OG
±15 psig ‡
.1
3500MMH2OG
±15PSIG
1
12FTH2O
–30INHG/15PSIG
.01
35KPAG
.01
±30INHGG
15 psia
Res
±400INH2OG
15PSIA
.01
±240ZING
30INHGA
.01
±1000GCMG
400INH2OA
.1
±760MMHGG
240ZINA
.1
±760TORRG
1000GCMA
±1000MBARG
1
760MMHGA
±1000CMH2OG
.1
760TORRA
±100KPAG
.1
1000MBARA
±0.1MPAG
1
1000CMH2OA
1
±1BARG
100KPAA
.1
±1KGCMG
0.1MPAA
.0001 ±1ATMG
1BARA
.001
30 psia
1KGCMA
.001 30PSIA
1ATMA
.001 60INHGA
15 psig vac ‡
Res
850INH2OA
15PSIVAC
480ZINA
.01
30INHGVAC
2100GCMA
.01
400INH2OVAC
1600MMHGA
.1
240ZINVAC
1600TORRA
.1
1000GCMVAC
1
2000MBARA
760MMHGVAC
.1
2100CMH2OA
760TORRVAC
.1
200KPAA
1000MBARVAC
1
0.2MPAA
1000CMH2OVAC
1
2BARA
C E CO MP
Div. of
Calibration
Zero button for gauge reference ranges
Pass code protected calibration via keypad
Non-interactive zero, span, and linearity, ±10% of range
Weight
Gauge:
9 ounces (approximately)
Shipping:
1 pound (approximately)
Materials
Standard: Extruded aluminum case, epoxy powder coated, ABS/
polycarbonate bezel, front and rear gaskets, polycarbonate label.
NEMA 4X: UV stabilized ABS/polycarbonate case, polycarbonate dis-
play window, polycarbonate front label, rear gasket, six stainless steel
cover screws. Not intended for permanent outdoor installations.
Connection and Material
1/4" NPT male fitting
All wetted parts are 316L stainless steel
Overpressure, Burst, Vacuum Service
3000 psig sensor: 5000 psig overpressure
5000 psig sensor: 7500 psig overpressure
All others:
Burst pressure:
Vacuum service:
Environmental Temperatures
Storage temperature:
Operating temperature:
Sensor compensated range:
Dimensions
See next page
‡ -HA option not available
Res
30 psia
Res
–15V100psig ‡
2KGCMA
.001 –15V100PSIG
.1
.0001 2ATMA
.001 –30INHG/100PSIG
.001
30 psig
Res
–30V200INHGG
.001 30PSIG
.01
–400V2770INH2OG
.001 60INHGG
.01
–240V1600ZING
Res
850INH2OG
1
–760V5200MMHGG 1
.01
480ZING
.1
–760V5200TORRG
.01
2100GCMG
1
–100V700KPAG
.1
1600MMHGG
1
–0.1V0.7MPAG
1600TORRG
–1V7BARG
.1
1
2000MBARG
–1V7KGCMG
1
1
2100CMH2OG
–1V7ATMG
.1
1
70FTH2O
100 psig
.1
.01
1
200KPAG
.1
100PSIG
1
0.2MPAG
.0001 200INHGG
.01
2BARG
.001 2770INH2OG
.1
2KGCMG
.001 1600ZING
.0001 2ATMG
.001 7000GCMG
60 psig
Res
.001
5200MMHGG
.001 60PSIG
5200TORRG
.01
.001 120INHGG
7000MBARG
.1
Res
1660INH2OG
7000CMH2OG
1
960ZING
230FTH2O
.01
1
4200GCMG
700KPAG
.01
1
.01
3100MMHGG
1
0.7MPAG
1
3100TORRG
1
7BARG
.1
4100MBARG
1
7KGCMG
1
4200CMH2OG
1
7ATMG
1
140FTH2O
.1
–15V200 psig ‡
1
400KPAG
.1
–15V200PSIG
0.4MPAG
.0001 –30INHG/200PSIG
1
4BARG
.001 –30V400INHGG
1
4KGCMG
.001 –400V5500INH2OG
.1
.0001 4ATMG
.001 –240V3200ZING
.001
100 psia
Res
–100V1400KPAG
.001 100PSIA
.1
–0.1V1.4MPAG
.001 200INHGA
.1
–1V14BARG
Res
2770INH2OA
1
–1V14KGCMG
.01
1600ZINA
1
–1V 14ATMG
200 psig
.01
7000GCMA
1
1
5200MMHGA
1
200PSIG
5200TORRA
400INHGG
.1
1
7000MBARA
5500INH2OG
1
1
7000CMH2OA
3200ZING
1
1
700KPAA
480FTH2O
1
.1
1
0.7MPAA
.0001 1400KPAG
1
7BARA
.001 1.4MPAG
.1
7KGCMA
.001 14BARG
.0001 7ATMA
.001 14KGCMG
.001
14ATMG
A
A
P
P
BSOLUTE
BSOLUTE
ROCESS
ROCESS
2 X pressure range overpressure
4 X sensor pressure rating, or 10,000 psi,
whichever is less
15 psia, 15 psig, ±15 psig, 30 psia, 100 psia,
100 psig, 200 psig
–40 to 203°F (–40 to 95°C)
–4 to 185°F (–20 to 85°C)
32 to 158°F (0 to 70°C)
Res
300 psig
Res
300PSIG
.1
.1
610INHGG
.1
.1
.1
4800ZING
1
1
700FTH2O
.1
1
2000KPAG
1
2MPAG
.001
1
20BARG
.01
1
20KGCMG
.01
.001 20ATMG
.01
500 psig
Res
.01
500PSIG
.01
.1
1020INHGG
.01
1
Res
1150FTH2O
1
.1
3500KPAG
1
.1
3.5MPAG
.001
1
35BARG
.01
1
35KGCMG
.01
1
35ATMG
.01
1000 psig
Res
1
1000PSIG
1
1
2040INHGG
1
1
2300FTH2O
1
1
7000KPAG
.1
1
7MPAG
.1
.0001
.0001 70BARG
.01
.001 70KGCMG
.01
.001 70ATMG
.01
.001
2000 psig
Res
Res
2000PSIG
1
.1
4070INHGG
1
4600FTH2O
.1
1
14MPAG
.1
.01
140BARG
1
.1
140KGCMG
1
.1
1
140ATMG
.1
.001
3000 psig
Res
.01
3000PSIG
1
.01
6100INHGG
1
.01
6900FTH2O
1
Res
20MPAG
.01
.1
200BARG
.1
200KGCMG
.1
.1
200ATMG
1
.1
5000 psig
Res
1
5000PSIG
.1
1
1
35MPAG
.01
.001 350BARG
.1
.01
350KGCMG
.1
.01
340ATMG
.1
.01
1220 American Way Libertyville, IL 60048
I
I
NSTRUMENTS
NSTRUMENTS
F20B, F22B
O ±0.25% Test Gauge Accuracy, ±0.1% Optional
O 316L Stainless Steel Wetted Parts
O Keypad Selectable Units and Auto Shutoff Times
O Store Readings in Memory
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
F20BN, F22BN
NEMA 4X
How to Specify
Housing Backlit Memory
F20B range - M4 - options
Standard
F20BBL range - M4 - options
F20BN range - M4 - options
NEMA 4X
F20BNBL range - M4 - options
F20B range - M8 - options
Standard
F20BBL range - M8 - options
F20BN range - M8 - options
NEMA 4X
F20BNBL range - M8 - options
F22B range - options
Standard
F22BBL range - options
F22BN range - options
NEMA 4X
F22BNBL range - options
Range—See table at left. Select a range code for default units.
psi = PSI
torr = TORR
inHg = INHG
mmH
O = MMH2O
2
oz/in
2
= ZIN
kg/cm
2
= KGCM
inH
O = INH2O
g/cm
2
= GCM
2
ftH
O = FTH2O
kPa = KPA
2
mmHg = MMHG
MPa = MPA
G
gauge reference pressure
A
VAC gauge reference vacuum
Options—add to end of model number. See price list for details.
HA
High accuracy, ±0.1% FS ±1 LSD. See range table.
PM
Panel mount, 4.1" x 4.1", n/a NEMA 4X
Sealed housing, CC for high-humidity food processing
FP
applications. NEMA models in absolute reference only.
MC
Metal front cover instead of plastic, n/a NEMA 4X
CS
Case bottom stiffener plate, n/a NEMA 4X
CC
Moisture resistant circuit board conformal coating
TP
Top port, gauge port on top of case, n/a NEMA 4X
SM
Surface mount plate, n/a NEMA 4X
CD
Calibration data, 5 test points and date
NC
NIST traceability documentation, 5 points and date
TP
Top gauge port. Primarily used with tire pressure
applications. Not available with NEMA 4X models.
Accessories—order separately
RB
High visibility orange rubber boot protects gauge
for portable applications. Not available with NEMA
4X models.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applications.
303SS body, 100 micron 304SS screen.
CON14SS
Quick connector to install or remove gauge with-
out tools. 304 stainless steel, urethane seal.
cecomp.com
800-942-0315
F20B, F22B
Quick Link
cecomp.com/bat
no
Yes
4 max
reading
no
Yes
no
Yes
8 max
reading
no
Yes
no
Yes
Min/max
no
Yes
mbar = MBAR
bar = BAR
cmH
O = CMH2O
2
atm = ATM
15V = –14.7 psi vac
30V = – 30 inHg vac
absolute reference
© 02-21

Advertisement

loading

Summary of Contents for Absolute Process Instruments Cecomp F20B

  • Page 1 Cecomp Battery Powered Digital Pressure Gauges w. Memory Options & Selectable Units F20B, F22B ® Ranges and Resolution Calibration O ±0.25% Test Gauge Accuracy, ±0.1% Optional See table below for standard ranges and units Zero button for gauge reference ranges O 316L Stainless Steel Wetted Parts Resolution is fixed for each engineering unit Pass code protected calibration via keypad...
  • Page 2 Installation Precautions, Ranges and Engineering Units, Basic Operation F20B, F22B Types of Gauges Bipolar sensors read positive pressure and vacuum in the same Absolute reference gauges read zero at full vacuum. With an open Gauge reference sensors always read zero with an open gauge port. units, and zero with the gauge port open.
  • Page 3 Basic Operation, Configuration F20B, F22B F20B Memory Gauge Configuration F22B Min/Max Setup With the gauge powered up and in the The gauge uses a 4 digit pass code to enter the configuration After the center power button is pressed normal operating mode, press and release modes.
  • Page 4 You may also return this product via pre-paid shipping stall the six screws. 50% of full-scale vacuum using the s and t buttons. to Absolute Process Instruments or your supplier for proper disposal. C E C OMP 1220 American Way Libertyville, IL 60048 cecomp.com...

This manual is also suitable for:

Cecomp f22bCecomp f20bblCecomp f20bnCecomp f20bnblCecomp f22bblCecomp f22bn ... Show all