Download Print this page

LIT 9000 Series Operating Manual

Lab oven

Advertisement

Quick Links

Version: August 2022
LIT 9000 series
Lab Oven
Operating Manual
Warranty
This Laboratory oven carries the New Zealand Industry standard 12 month return-to-base warranty
It does not cover misuse of the oven or failure to follow the user instructions in this manual
L.I.T. Lab ovens must be kept in an environment with < 50% RH to avoid corrosion
Laboratory & Industrial Technologies
1

Advertisement

loading

Summary of Contents for LIT 9000 Series

  • Page 1 Version: August 2022 LIT 9000 series Lab Oven Operating Manual Warranty This Laboratory oven carries the New Zealand Industry standard 12 month return-to-base warranty It does not cover misuse of the oven or failure to follow the user instructions in this manual L.I.T.
  • Page 2: Technical Specifications

    CCC / CE certified 9000 series laboratory ovens 220vac to 240vac, 50hz, 15A, single phase. Electric power requirements: RT+10C to 250C Temperature range: 0 to 9999 Minutes (166.65 Hrs) Timer range: Technical Specifications Model LIT-9050 LIT-9140 LIT-9240 LIT-9420 400L Internal volume litres...
  • Page 3 PCD E9000 Controller Operating instructions 1. Operation 1) When the controller is switched on, the display illuminates all segments for 3 seconds, then briefly flashes a parameter screen then it starts running the current set parameters. Press the SET button to set the temperature and then again to set the time 2) When the controller is in the temperature setting mode, the lower window displays the temperature set- point “SP”...
  • Page 4 PCD E9000 Controller Auto-tuning the controller's internal PID Only use the auto-tuning function if there is excessive temperature overshoot ! Press the AT button for 5 seconds. then the auto-tune “AT” indicator flashes, after auto-tune ends, the indicator stops flashing, and the new AT parameter value is saved automatically. To enable A/T In stop mode, press the “AT”...
  • Page 5: Controller Wiring

    Internal parameter settings Press the “Set” button for 3 seconds, controller will display the password prompt “Lc”. If you want to adjust the password, then press the “Set” button again, it will then go into the internal parameter setting state. If you press the “Set” button for another 3 seconds, it returns to the run state. Parameter list: Parameter (Setting range)
  • Page 6: Electrical Safety

    Electrical Safety To prevent FIRE or SHOCK hazard, DO NOT expose this product to any moisture There is a 10A safety circuit breaker on the front panel next to the on/off button This appliance must be periodically tested according to AS/NZS 3760 ...